How AI should be used in radiology: assessing ambiguity and completeness of intended use statements of commercial AI products

Abstract

Background

Intended use statements (IUSs) are mandatory to obtain regulatory clearance for artificial intelligence (AI)-based medical devices in the European Union. In order to guide the safe use of AI-based medical devices, IUSs need to contain comprehensive and understandable information. This study analyzes the IUSs of CE-marked AI products listed on AIforRadiology.com for ambiguity and completeness.

Methods

We retrieved 157 IUSs of CE-marked AI products listed on AIforRadiology.com in September 2022. Duplicate products (n = 1), discontinued products (n = 3), and duplicate statements (n = 14) were excluded. The resulting IUSs were assessed for the presence of 6 items: medical indication, part of the body, patient population, user profile, use environment, and operating principle. Disclaimers, defined as contra-indications or warnings in the IUS, were identified and compared with claims.

Results

Of 139 AI products, the majority (n = 78) of IUSs mentioned 3 or less items. IUSs of only 7 products mentioned all 6 items. The intended body part (n = 115) and the operating principle (n = 116) were the most frequently mentioned components, while the intended use environment (n = 24) and intended patient population (n = 29) were mentioned less frequently. Fifty-six statements contained disclaimers that conflicted with the claims in 13 cases.

Conclusion

The majority of IUSs of CE-marked AI-based medical devices lack substantial information and, in few cases, contradict the claims of the product.

Critical relevance statement

To ensure correct usage and to avoid off-label use or foreseeable misuse of AI-based medical devices in radiology, manufacturers are encouraged to provide more comprehensive and less ambiguous intended use statements.

Key points

• Radiologists must know AI products’ intended use to avoid off-label use or misuse.

• Ninety-five percent (n = 132/139) of the intended use statements analyzed were incomplete.

• Nine percent (n = 13) of the intended use statements held disclaimers contradicting the claim of the AI product.

• Manufacturers and regulatory bodies must ensure that intended use statements are comprehensive.

Graphical Abstract

Introduction

Artificial intelligence (AI) products are entering radiology departments with already over 200 CE-marked commercial products available for use in the European Union (EU) [1, 2]. In order to help the radiologist in routine clinical practice, these AI products are designed to help with, e.g., segmentation of healthy tissue and lesions, detection of critical findings and consecutive patient triage, or differential diagnosis of certain disease states [3, 4].

Medical AI applications are considered software as a medical device (SaMD) according to EU legislation and therefore have to comply with the EU Medical Device Regulation (MDR) in order to be marketed and clinically used. The so-called intended use statement (IUS) of SaMD is an important step in the process of compliance and obtaining regulatory clearance, such as a CE mark or FDA clearance [5,6,7,8].

In the MDR, the terms intended use and intended purpose are used interchangeably and defined as follows: ‘intended purpose’ means the use for which a device is intended according to the data supplied by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements and as specified by the manufacturer in the clinical evaluation (Chapter 1, 2.12). However, the exact elements that should be part of this intended use statement are not clearly stated in the above-cited MDR definition of the IUS itself but can be found elsewhere in the MDR’s technical documentation (Annex II, 1.1.) [9]. Additional documentation by the International Organization for Standardization (ISO) and the International Electrotechnical Commission (IEC) provides more guidance. For example, ISO 14971:2019 for the application of risk management to medical devices states: “the intended use should take into account information such as the intended medical indication, patient population, part of the body or type of tissue interacted with, user profile, use environment, and operating principle” [10,11,12,13].

Unfortunately, as of now, the IUS of CE-marked medical AI products is rarely made publicly available by manufacturers, making it difficult for (potential) users to track down the correct use of their medical device in clinical practice. A critical discussion of intended use statements by radiological societies is hampered by this lack of information. Understanding the intended use of AI products by radiologists is critical to the safe operation of the device, and consequently for patient safety, as underlined by MDR I (43) with the following statement: “Transparency and adequate access to information, appropriately presented for the intended user, are essential in the public interest, to protect public health, to empower patients and healthcare professionals and to enable them to make informed decisions.” Therefore, IUSs of medical AI products need to be comprehensive and clear, without ambiguity. However, this has not been analyzed systematically, yet.

To fill this gap, this study aims to investigate the completeness and potential ambiguity of IUSs for AI products in radiology with respect to the requirements made within the MDR and ISO/IEC standards. Additionally, we make the statements publicly available on www.AIforRadiology.com to increase transparency.

Methods

Product inclusion

Inclusion was based on product listings on AIforRadiology.com. This platform maintains a database of CE-marked AI products for radiology. In March 2022, all vendors with listed products (87 vendors, 191 products) were requested to submit or verify an IUS for their product(s). Any new products added to the listing were included until September 26, 2022 (n = 12), leading to a total of 203 products. For 46 AI products, the vendor did not respond; therefore, these products were not analyzed. In total, 157 IUSs were collected. Discontinued products (n = 3) and duplicate products (n = 1) were removed from the analysis. One vendor (Quibim) had 15 biomarker-based products listed with the same intended use. To prevent overrepresentation of this single IUS in the analysis, we included only one product, and the other 14 were excluded. In total, 139 products and their respective IUSs were eligible for analysis. The full list of evaluated AI products is available in the Supplementary materials.

For each product, the CE risk class (I, IIa, IIb, or III) at that time was documented as well as the relevant regulation under which it was certified, either the former EU Medical Device Directive (MDD) (pre-May 2021) or current EU MDR. Products were also classified according to their main function or task (quantification, detection, diagnosis, triage, image enhancement).

Intended use statement items

Since no clear definition of essential intended use statement items is given in the MDR, we scrutinized the intended use statement for the presence (yes/no) of the following six intended use statement items as described in several relevant ISO/IEC standards [10,11,12,13]:

  • Intended medical indication

  • Intended part of the body or type of tissue applied to or interacted with

  • Intended patient population

  • Intended user profile

  • Intended use environment

  • Intended operating principle

Further description of the IUS items is available in Table 1. First, the authors (K.G.v.L., D.M.H., S.S.) scored the same IUS of 30 products, after which a meeting was held to align our definitions of the IUS items and to obtain consensus in case of discrepant scoring. The remaining IUS were scored by one of the authors (K.G.v.L., D.M.H., S.S.), after which a consensus meeting was held to discuss uncertainties in scoring.

Table 1 Description of defined item derived from the intended use statements

Furthermore, it was documented if contra-indications or warnings were made in the IUS. These items were considered disclaimers. Disclaimer statements included wording such as “act per the standard of care” and “not for diagnostic use.” The disclaimers were compared with claims in the IUS, if present. The presence of potential conflicts of claims and disclaimers were categorized as yes/no.

Analysis

The overall completeness of the IUSs of the AI products was assessed by counting the presence of the IUS items and expressing this as a percentage of the total number of products analyzed. A subanalysis was subsequently performed based on the regulatory risk classification and the main intended task of the product. Similarly, the conflicting claims and disclaimers were counted. A histogram was constructed to show the level of completeness of each product.

IUSs for which each respective vendor agreed to make publicly available can be found on www.AIforRadiology.com.

Results

Product inclusion

We received an IUS for 157 out of 203 products, of which 139 were included for analysis.

The included products were certified under the current MDR (class IIa = 23, class IIb = 9) and former MDD (class I = 52, class IIa = 55). Most products were intended for quantification (n = 54), detection (n = 36), or diagnosis (n = 25). A smaller number of products focused on triage (n = 18) or image enhancement (n = 6).

Completeness based on intended use statement items

Only 7 products had an IUS that contained all of the 6 defined items. The majority (n = 78) mentioned 3 or less items (0 items, n = 1; 1 item, n = 18; 2 items, n = 21; 3 items, n = 38). Four items were mentioned in 32 statements and 5 items in 22 statements (Fig. 1).

Fig. 1
figure 1

Histogram of the number of items present per intended use statement ranging from 0 to 6 items. Intended use items scored were intended medical indication, intended part of the body or type of tissue applied to or interacted with, intended patient population, intended user profile, intended use environment, and intended operating principle

Table 2 gives the number of instances each item was present in the IUS. The intended body part (n = 115) and the operating principle (n = 116) were most often mentioned. The intended use environment (n = 24) and the intended population (n = 29) were only mentioned scarcely (Fig. 2). Deidentified example statements of good and poor IUSs are given in Table 3.

Table 2 Overview of the specific intended use items for all products and per product type and level of certification
Fig. 2
figure 2

Two spider charts of the percentage of IUS items present per regulatory class of the product (a) and product type (b). MDD, Medical Devices Directive; MDR, Medical Devices Regulation

Table 3 Examples of two intended use statements which had a score of 6/6 items present (complete) and two examples that had a score of maximal 2/6 items present (incomplete)

Ambiguity based on conflicting claims and disclaimers

Of the 139 included statements, 56 held disclaimers. For 13 of these products, the claims and disclaimers were flagged to contradict each other. An overview of the number of disclaimers per product type and regulatory class is shown in Fig. 3. Potential discrepant disclaimer statements included, for example, “act per the standard of care” (n = 7) and “not for diagnostic use” (n = 6), while claiming to aid in the diagnosis, triaging, or risk scoring of clinical conditions. Example disclaimer statements are available in Table 4.

Fig. 3
figure 3

Bar chart of the number of intended use statements without disclaimers (blue), with disclaimers that did not interfere with the claim of the product (orange), and disclaimers that contradict the claim of the product (gray), per regulatory class and product type. MDD, Medical Devices Directive; MDR, Medical Devices Regulation

Table 4 Two examples of disclaimers that contradicted the claim in the intended use statement

Discussion

Many AI products in radiology have an ambiguous or incomplete IUS, as provided to AIforRadiology. Of 139 analyzed products, 78 (56%) were lacking at least 3 essential items in their IUS. Furthermore, 13 (9%) of IUSs held disclaimers that contradicted the claims of the product. The IUS is an important aspect of regulatory documentation as it defines what is considered safe and proper use of the medical device. This is also expressed by the frequent use in the MDR: “Intended purpose” is mentioned a staggering 87 times and “intended use” 31 times. Moreover, the clinical evidence needed to be granted a CE mark is based on the intended use as described in the IUS. Also, it restricts the use of the product by the intended user(s) outside of its intended scope. Therefore, users should be aware that the IUSs of AI products, as provided for this study, were often incomplete and that additional information may be needed to correctly use such AI products in clinical practice.

The intended patient population and use environment were not present in most IUSs. The next most common was a lack of information on the medical indication and the user profile. It is not clear why the intended patient population was only present in 29 of the 139 IUSs. It seems obvious that the intended patient population of an AI product is of utmost importance for safe use in clinical practice, especially since AI systems are prone to bias when trained on a non-representative population. The use environment was only reported in 24 of 139 IUSs and maybe a more difficult item to translate to AI products. However, users should be informed in which (clinical) situation the product is applicable. For instance, an AI product that is designed for screening situations may not be safe to use in a clinical environment.

Our findings are acknowledged by the Medicines & Healthcare products Regulatory Agency (MHRA). This agency mentioned vague intended purposes as a “common issue” and has recently published a guideline to craft IUSs to mitigate this issue for the UK market [14]. These recommendations encourage mentioning explicit items in the IUS that were outlined and used for analysis in this study. We would recommend vendors and regulatory consultants to include all defined IUS items in their public documentation to provide transparent information and to ensure the safe use of AI products in clinical practice, as required by the MDR.

In some of the analyzed statements, we found relevant contra-indications or disclaimers, which are important means to avoid unsafe use, going beyond the actual intended use of the product as a medical device. However, if there is a direct contradiction between the “perceived” intended use of a medical device, e.g., by claims made in the marketing material or suggested in the name of a medical device, disclaimers can add to the confusion about the proper use of a medical device. It is a requirement of all global medical device regulations that marketing material is consistent with regulatory-approved claims and that a device’s labeling covers all externally available marketing material, including manufacturer websites, brochures, and even verbal communications with customers. A recent study already found that 13% of the products were marketed as AI or machine learning, while this was not reflected in the US Food and Drug Administration (FDA) 510 (k) clearance [15]. Further research should focus on how well the intended use in regulatory documents aligns with marketing material and whether disagreement between those two may be a potential source of patient harm.

A clear understanding of the intended use of an AI product is crucial to mitigate the risk of both off-label use and foreseeable misuse. Both vendors and users share the responsibility for the correct and appropriate use of the product. Vendors must take proactive measures to minimize the risk of foreseeable misuse and address potential dangers. On the other hand, users must understand the intended use of the medical device they purchase. Because users, typically radiologists, may be held liable for civil claims and damages in case of any adverse events resulting from the AI product’s misuse [5]. This was made explicit in the example from the USA from April 2022 when the FDA sent a letter to healthcare providers as it became clear that AI products for stroke triage were not used as intended. In a recent statement to healthcare providers, the FDA pointed out that AI-based software tools for the detection of large vessel occlusion in stroke patients need to be used strictly according to their intended use in order to avoid misdiagnosis resulting in patient injury or death [16].

Our study has two main limitations. The first is the subjective nature of analyzing descriptive text. The interpretation of certain IUS elements can be open to debate. Therefore, we tried to be conservative in our analysis. If one of the items was mentioned, but was not entirely clearly stated, we still classified such an item as present. Vague statements were discussed in a panel to overcome inter-reader variability and minimize personal bias. Secondly, the absence of a comprehensive public database containing the IUSs of CE-marked AI products in radiology posed a challenge. Consequently, we had to rely on the information provided by vendors to AIforRadiology. Unfortunately, not all vendors supplied an IUS for their products upon request. Some may have shared a more minimal version of their IUS causing some of the elements to be missing according to our analysis. The European Commission has been developing a publicly accessible database called EUDAMED, in which registering medical devices with complete information will become mandatory by 26 May 2024. However, EUDAMED registration is currently voluntary, and therefore, IUSs for all devices are not yet available.

Conclusion

In conclusion, our analysis revealed a significant number of AI products listed on AIforRadiology.com with incomplete or ambiguous IUSs. These omissions and/or vague statements, increase the risk of off-label or erroneous application of AI in clinical practice. It is imperative that regulators and central bodies offer comprehensive guidance for the IUS of medical devices including medical AI. Until the enforcement of the EUDAMED database, we encourage manufacturers to be transparent with their regulatory-approved intended use, accurately reflected in all marketing. In addition, we encourage radiologists to get familiar with the intended use, including all its elements, of the AI products they employ, to ensure safe patient care.

Availability of data and materials

The datasets generated during and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

AI:

Artificial intelligence

CE:

Conformité Européene (European Union’s mandatory conformity marking)

FDA:

US Food and Drug Administration

IEC:

International Electrotechnical Commission

ISO:

International Organization for Standardization

IUS:

Intended use statement

MDD:

Medical Devices Directive

MDR:

Medical Devices Regulation

References

  1. van Leeuwen KG, Schalekamp S, Rutten MJCM, van Ginneken B, de Rooij M (2021) Artificial intelligence in radiology: 100 commercially available products and their scientific evidence. Eur Radiol. https://doi.org/10.1007/s00330-021-07892-z

    Article  PubMed  PubMed Central  Google Scholar 

  2. Diagnostic Image Analysis Group R (2020) AI for radiology: an implementation guide. Available via www.AIforRadiology.com. Accessed 26 Sep 2022

  3. Topol EJ (2019) High-performance medicine: the convergence of human and artificial intelligence. Nat Med 25:44–56

    Article  CAS  PubMed  Google Scholar 

  4. van Leeuwen KG, de Rooij M, Schalekamp S, van Ginneken B, Rutten MJCM (2022) How does artificial intelligence in radiology improve efficiency and health outcomes? Pediatr Radiol 52:2087–2093

    Article  PubMed  Google Scholar 

  5. Hedderich DM, Weisstanner C, Van Cauter S et al (2023) Artificial intelligence tools in clinical neuroradiology: essential medico-legal aspects. Neuroradiology 65:1091–1099

    Article  PubMed  PubMed Central  Google Scholar 

  6. Muehlematter UJ, Daniore P, Vokinger KN (2021) Approval of artificial intelligence and machine learning-based medical devices in the USA and Europe (2015–20): a comparative analysis. Lancet Digit Health 3:e195–e203

    Article  CAS  PubMed  Google Scholar 

  7. Nelson BJ, Zeng R, Sammer MBK, Frush DP, Delfino JG (2023) An FDA guide on indications for use and device reporting of artificial intelligence-enabled devices: significance for pediatric use. J Am Coll Radiol 20:738–741

    Article  PubMed  Google Scholar 

  8. Petrick N, Chen W, Delfino JG et al (2023) Regulatory considerations for medical imaging AI/ML devices in the United States: concepts and challenges. J Med Imaging (Bellingham) 10:051804

    PubMed  Google Scholar 

  9. Regulation (EU) 2017/745 of the European parliament and of the council (2017) On Medical Devices, Amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and Repealing Council Directives 90/385/EEC and 93/42/EEC. Retrieved from: https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32017R0745.

  10. The International Organization for Standardization (2019) 14971: Medical devices — Application of risk management to medical devices (Vol. 14971). Retrieved from: https://www.iso.org/obp/ui/#iso:std:iso:14971:ed-3:v1:en.

  11. The International Organization for Standardization (2015) 62366-1: Medical devices — Part 1: Application of usability engineering to medical devices. Retrieved from: https://www.iso.org/obp/ui/en/#iso:std:iec:62366:-1:ed-1:v1:en.

  12. The International Organization for Standardization (2016) 62366-2: Medical devices — Part 2: Guidance on the application of usability engineering to medical devices. Retrieved from: https://www.iso.org/obp/ui/#iso:std:iec:tr:62366:-2:ed-1:v1:en.

  13. The International Organization for Standardization (2016) 82304-1: Health software — Part 1: General requirements for product safety. Retrieved from: https://www.iso.org/obp/ui/en/#iso:std:iec:82304:-1:ed-1:v1:en.

  14. Medicines & Healthcare products regulatory agency (2023) Crafting an intended purpose in the context of software as a medical device (SaMD). Retrieved from: https://www.gov.uk/government/publications/crafting-anintended-purpose-in-the-context-of-software-as-a-medical-device-samd/crafting-an-intended-purpose-in-thecontext-of-software-as-a-medical-device-samd.

  15. Clark P, Kim J, Aphinyanaphongs Y (2023) Marketing and US Food and Drug Administration clearance of artificial intelligence and machine learning enabled software in and as medical devices: a systematic review. JAMA Netw Open 6:e2321792

    Article  PubMed  PubMed Central  Google Scholar 

  16. U.S. Food and Drug Administration (2023) Intended use of imaging software for intracranial large vessel occlusion - letter to health care providers

    Google Scholar 

Download references

Funding

No funding.

Author information

Authors and Affiliations

Authors

Contributions

All authors read and approved the final manuscript.

Corresponding author

Correspondence to Steven Schalekamp.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

DMH, SS, and KGvL declare that they have no competing interests.

HH: Hardian Limited.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

List of evaluated AI products.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

van Leeuwen, K.G., Hedderich, D.M., Harvey, H. et al. How AI should be used in radiology: assessing ambiguity and completeness of intended use statements of commercial AI products. Insights Imaging 15, 51 (2024). https://doi.org/10.1186/s13244-024-01616-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13244-024-01616-9

Keywords