Performance comparison of ultrasonography and magnetic resonance imaging in their diagnostic accuracy of placenta accreta spectrum disorders: a systematic review and meta-analysis

Abstract

Objectives

Accurate prenatal diagnosis of placenta accrete spectrum disorder (PAS) remains a challenge, and the reported diagnostic value of ultrasonography (US) and magnetic resonance imaging (MRI) varies widely. This study aims to systematically evaluate the diagnostic accuracy of US as compared with MRI in the detection of PAS within the identical patient population.

Methods

Medline, EMBASE, Google scholar and Cochrane library were searched. Pooled sensitivity, specificity, diagnostic odds ratio (DOR) and the area under the summary receiver operating characteristic (SROC) curve were calculated. Subgroup analysis was also performed to elucidate the heterogeneity of results.

Results

A total of 18 articles comprising 861 pregnancies were included in the study. The overall diagnostic accuracy of US for identification of PAS was as follows: sensitivity [0.90 (0.86–0.93)], specificity [0.83 (0.79–0.86)], DOR [39.5 (19.6–79.7)]. The overall diagnostic accuracy of MRI for identification of PAS was as follows: sensitivity [0.89 (0.85-0.92)], specificity [0.87 (0.83–0.89)], DOR [37.4 (17.0–82.3)]. The pooled sensitivity (p = 0.808) and specificity (p = 0.413) between US and MRI are not significantly different. SROC analysis revealed that there was no statistical difference (p = 0.552) in US and MRI for the overall predictive accuracy of PAS. Furthermore, in the subgroup analysis of between retrospective and prospective studies, between earlier and most recent studies, there was no statistical difference (p > 0.05) in diagnostic accuracy of US and MRI for the detection of PAS.

Conclusions

Both ultrasonography (US) and magnetic resonance imaging (MRI) showed comparable accuracy in the prenatal diagnosis of placenta accrete spectrum disorder (PAS). Routine employment of MRI with relatively high expense in the prenatal identification of PAS should not be recommended.

Key points

  • Both US and MRI showed comparable accuracy in the prenatal diagnosis of PAS.

  • MRI should not be used routinely as an adjunct tool to US in the diagnosis of PAS.

  • The comprehensive scoring system using US, MRI and clinical risk factors can help to improve accurate prenatal diagnosis of PAS.

Introduction

Placenta accrete spectrum (PAS) disorder is a growing obstetric concern. The rise in cesarean section rates worldwide over the past decades has led to a fundamental increase in the prevalence of PAS [1]. PAS can be classified into three major variants of placenta accreta, placenta increta and placenta percreta according to the degree of trophoblastic invasion through the myometrium and the uterine serosa [2].

Various predisposing factors have been identified for the development of PAS; this includes placenta previa, prior cesarean delivery, multiparity, advanced maternal age (> 35 years), myometrial scaring from prior uterine surgery, and assisted reproductive techniques. Of which, the combination of placenta previa and previous cesarean delivery is the most commonly reported risk factor. PAS leads to the formation of hypertrophied, disorganized uteroplacental vascularity at the implantation site, and the immediate clinical consequence is massive hemorrhage while removing the placenta manually after delivery. PAS is the most common indication for emergent peripartum hysterectomy [3, 4] and also leads to extensive morbidities (i.e., massive transfusion, disseminated intravascular coagulation, infection, sepsis) and surgical-related complications (i.e., pelvic organs injury, fistula formation) [5].

Prenatal diagnosis of PAS allows for meticulous planning and multidisciplinary management of patients, thus substantially improve the maternal prognosis. However, accurate identification of morbidly adherent placenta before delivery remains a challenge, and almost half to two-thirds of PAS remain undiagnosed before delivery indicated by recent studies [6, 7]. Ultrasonography (US) is the most commonly used diagnostic modality for detecting abnormal placentation, while magnetic resonance imaging (MRI) is usually recommended as an adjunctive diagnostic modality and reserved for cases with inconclusive ultrasound findings. Although the diagnostic values of US and MRI has been previously studied, there is a high variability in the reported diagnostic accuracy, which can be attributed by the relative rarity of the disease and the subjectivity of the imaging modalities largely dependence on the experiences of the sonographer or radiologist.

Therefore, this study was performed to evaluate and compare the diagnostic performance of the two imaging modalities (US and MRI) for the prenatal identification of PAS. To our knowledge, this study is the first systemic review and meta-analysis comparing the predictive accuracy between MRI and US in PAS within the identical patient population.

Materials and methods

Search strategy

A systemic literature search using Medline, Embase, Google Scholar, the Cochrane Database of Systematic Reviews (CDSR), the Cochrane Central Register of Controlled Trials (CENTRAL), and Database of Abstracts of Reviews of Effects (DARE) up to May 28, 2021, using the relevant keywords: “magnetic resonance imaging,” “ultrasound,” “sonography,” “ultrasonography,” “placenta accrete,” “invasive placenta,” “placenta increta,” as well as combinations of these terms. Two authors independently reviewed the abstracts and full-text; relevant data were collected and analyzed.

Study selection

Two independent investigators reviewed the literature and performed the study selection. Inconsistencies were discussed among the reviewers, and consensus was reached.

Inclusion criteria for this meta-analysis were as follows:

  1. (1)

    Confirmation of PAS defined as histopathological evidence and/or intraoperative finding of gross placenta invasion.

  2. (2)

    Prospective cohorts and retrospective cohorts studies evaluating the predictive value of both US and MRI within the same patient population.

  3. (3)

    Presence of US signs indicative of PAS, using the reference findings as follows: prominent lacunae, myometrial thinning, focal exophytic masses, bladder line interruption, increased vascularity.

  4. (4)

    Presence of MRI signs indicative of PAS, using the reference findings as follows: intraplacental dark bands with a random pattern on T2 weighted sequences, heterogeneous signal intensity in the placenta, focal uterine bulge, focal interruptions of the uterine wall, and tenting of the bladder.

Exclusion criteria for this meta-analysis were as follows:

  1. (1)

    Opinions, editorials, review articles, case reports, and studies without the necessary variables mentioned above.

  2. (2)

    Absence of a precise number of participants, true positives (TP), false positives (FP), false negatives (FN), and true negatives (TN) of MRI and US assessment or such data cannot be calculated.

Data extraction

Two authors independently collected data based on the inclusion criteria and exclusion criteria. The following information was obtained and listed from the included study: year of publication, first author’s name and country, number of patients, number of PAS, patient characteristics, study design (prospective or retrospective), technique aspects, image interpretation (blind or not). The absolute number of patients and relevant data including TP, FP, TN, and FN for US and MRI were collected from each included study. We calculated the above data if such information was not directly reported. We also contacted the authors for any missing data for the completion of data extraction.

Quality Assessment

The updated QUADAS-2 (Quality Assessment of Diagnostic Accuracy Studies-2) criteria were employed to evaluate the risk of bias and applicability of the included studies. Each item was rated as “low,” “high,” or “unclear.” We scored the item as “unclear” when the absolute information was absent to conclude an accurate judgment.

Statistical analysis

Meta-Disc software for meta-analysis was used for statistical analyses [8]. Statistical difference was defined as p < 0.05.

The heterogeneity of studies included was evaluated by I2 statistic and Cochran’s Q test. The study was defined as insignificant heterogeneity with I2 values of 0–25%, low heterogeneity with I2 values of 25–50%, moderate heterogeneity with I2 values of 50–75%, and high heterogeneity with I2 values of 75–100%. To calculate the overall diagnostic accuracy of US and MRI for the identification of PAS including pooled sensitivity and specificity, positive and negative likelihood ratio (LR+ and LR−), and diagnostic odds ratio (DOR), appropriate statistical modality was chosen according to the outcomes of heterogeneity test: a random effect model used when I2 exceeded 50%, a fixed-effect model used in other cases.

For comparison of diagnostic performance between US and MRI, the hierarchical summary receiver operating characteristics (SROC) curves were plotted. The area under the SROC curve (AUC) and the Q* index (the point in SROC curve space closest to the ideal top-left corner and at which test sensitivity and specificity are equal) were also computed. To determine the difference of sensitivity, specificity, and AUS between US and MRI, Z value tests were performed. Additionally, sensitivity analysis was performed to assess the effect of each included study on the overall results.

Results

Literature search and general characteristics of the included studies

A total of 531 primary citations were identified from the comprehensive literature search. Among them, 455 studies were excluded based on screening the titles and abstracts. Two investigators carefully read the full-text manuscripts of the remaining 76 studies, and 58 studies were excluded according to the inclusion criteria. Finally, 18 studies (Levine [9], Masselli [10], Dwyer [11], Mansour [12], Lim [13], Elhawary [14], Peker [15], Riteau [16], Algebally [17], Satija [18], Balcacer [19], Rezk [20], Marcillac [21], Budorick [22], Kumar [23], Ayati [24], Romeo [25], Xia [26]) were included in our meta-analysis (Fig. 1). There is a total of 861 pregnancies in the included studies with a potential risk of PAS, and 339 (39.4%) pregnancies were diagnosed with PAS based on the histopathological evidence and/or intraoperative findings of gross placenta invasion. Of the included 18 studies, the majority of them (83.3%) specifically indicated that both the radiologists and sonographers reviewed and interpreted the images without prior knowledge of the pathological and surgical findings to avoid observer bias. The general characteristics of the included studies are summarized in Table 1.

Fig. 1
figure 1

Flowchart of studies for meta-analysis

Table 1 General characteristics of studies included in the systematic review (18 studies)

A quality assessment of the included 18 studies by the updated QUADAS-2 is outlined in Fig. 2. The results indicated an overall satisfactory quality of the identified studies. Furthermore, sensitivity analysis was conducted to assess the effect of each included study on the overall results by omitting one study at a time. No outlier was identified.

Fig. 2
figure 2

Summary results of QUADAS tool on risk of bias and applicability concerns for the included studies in the present systematic review

Overall diagnostic accuracy and comparison between US and MRI for the prenatal identification of PAS

The overall diagnostic accuracy of US for identification of PAS was as follows: sensitivity [0.90 (0.86–0.93)] (Fig. 3A), specificity [0.83 (0.79–0.86)] (Fig. 3B), positive likelihood ratios (LR +) [4.85 (3.08–7.64)], negative likelihood ratios (LR-) [0.16 (0.12–0.23)], DOR [39.5 (19.6–79.7)] (Fig. 4A). The overall diagnostic accuracy of MRI for identification of PAS was as follows: sensitivity [0.89 (0.85–0.92)] (Fig. 5A), specificity [0.87 (0.83–0.89)] (Fig. 5B), LR+ [5.16 (3.42–7.78)], LR− [0.17 (0.11–0.27)], DOR [37.4 (17.0–82.3)] (Fig. 6). The pooled diagnostic sensitivity (p = 0.808) and specificity (p = 0.413) between US and MRI are not significantly different in the detection of PAS.

Fig. 3
figure 3

Forest plots of the overall sensitivity (A) and specificity (B) of diagnosis of placenta accrete spectrum disorder (PAS) with ultrasonography (US). The first author’s name of each study is listed. CI, confidence interval

Fig. 4
figure 4

The diagnostic performance of ultrasonography (US) for the prenatal identification of placenta accrete spectrum disorder (PAS) based on the (A) forest plots of overall diagnostic odds ratio (DOR) and (B) summary receiver operating characteristic (SROC) curve. The weight of each study is reflected by the size of data points. AUC, area under the curve; SE, standard error

Fig. 5
figure 5

Forest plots of the overall sensitivity (A) and specificity (B) of diagnosis of placenta accrete spectrum disorder (PAS) with magnetic resonance imaging (MRI). The first author’s name of each study is listed. CI, confidence interval

Fig. 6
figure 6

The diagnostic performance of magnetic resonance imaging (MRI) for the prenatal identification of placenta accrete spectrum disorder (PAS) based on the A forest plots of overall diagnostic odds ratio (DOR) and B summary receiver operating characteristic (SROC) curve. The weight of each study is reflected by the size of data points. AUC, area under the curve; SE, standard error

The SROC curve for the performance of prenatal diagnosis of US and MRI for the identification of PAS are shown in Figs. 4B and 6B, respectively. Based on the SROC curve of US and MRI, we calculated the area under the curve (AUC) and Q* indices for predictive accuracy and the results show that AUC and Q* indices of US were 0.935 (standard error (SE), 0.017) and 0.872, that of MRI was 0.932 (SE 0.021) and 0.867, respectively. Z value statistics analysis was used to compare the AUC between US and MRI with a value of 0.1299, and there was no statistical significance with a corresponding p value of 0.5517.

Subgroup analysis of US and MRI in the prenatal identification of PAS

Given that the predictive accuracy of imaging is evolving and can be influenced by either technique improvement or accumulative experience of sonographers and radiologists. To investigate the heterogeneity of identified 18 studies, we conducted two subgroup analyses by dividing the included studies into (1) retrospective studies versus prospective studies, (2) early studies (published before 2015) versus recent studies (published after 2015).

In the subgroup analysis of prospective studies [9, 10, 12, 17, 18, 20, 23, 24], the pooled performance of US was sensitivity 92%, specificity 86%, AUC 0.96 (SE 0.01), and that of MRI was sensitivity 93%, specificity 91%, AUC 0.96 (SE 0.02). In retrospective studies [11, 13,14,15,16, 19, 21, 22, 25, 26], the pooled performance of US was sensitivity 88%, specificity 80%, AUC 0.96 (SE 0.03) and that of MRI was sensitivity 85%, specificity 82%, AUC 0.89 (SE 0.04).

In the subgroup analysis of early studies (published before 2015) [9,10,11,12,13,14,15,16,17], the pooled performance of US was sensitivity 90%, specificity 88%, AUC 0.94 (SE 0.02), and that of MRI was sensitivity 90%, specificity 89%, AUC 0.97 (SE 0.01). In recent studies (published after 2015) [18,19,20,21,22,23,24,25,26], the pooled performance of US was sensitivity 90%, specificity 79%, AUC 0.91 (SE 0.04) and that of MRI was sensitivity 88%, specificity 85%, AUC 0.90 (SE 0.02).

There was no statistical significance (p > 0.05) in pooled sensitivity and specificity as well as AUC of SROC curves between US and MRI in different subgroups. The summary estimates of sensitivity, specificity, DOR, LR+, LR−, SROC curve (AUC ± SE, Q* ± SE) of US and MRI for prenatal identification of PAS in the subgroup of retrospective and prospective studies, and the subgroup of earlier and recent studies are listed in Table 2, respectively.

Table 2 Summary estimates of sensitivity, specificity, DOR, LR+, LR−, SROC curve (AUC ± SE, Q* ± SE) of US and MRI for identification of PAS in different subgroup

Discussion

Since the first study describing the ultrasound characteristics of placenta increta in the early 1980s [27], ultrasonography (US) has been used as the primary diagnostic tool for PAS. US is a widely available, inexpensive diagnostic tool and is generally considered safe for both mother and fetus. However, the diagnostic accuracy of US for identification of invasive placenta could be impaired by several clinical parameters such as unfavorable placenta site (i.e., posterior placenta) and high body mass index (BMI). Accurate prenatal diagnosis of PAS remains a challenge.

On the contrary, the imaging modality of MRI is not restricted by the depth and it offers the advantage of multiplanar imaging and excellent soft-tissue resolution. MRI demonstrates an excellent performance for defining the topography and depth of abnormal placentation [10]. MRI is often recommended when the adjacent pelvic organs involvement is highly suspected, or there is an inconclusive assessment on US, or the posterior placenta and maternal obesity affect the diagnostic accuracy of US. A recent systematic review and meta-analysis [28] was conducted to evaluate the predictive performance of MRI in detecting the severity of abnormal placental invasion and showed that the pooled sensitivity of MRI in detecting the placenta accrete was 94.4%, 100% for placenta increta, 86.5% for placenta percreta, and corresponding pooled specificity was 98.8%, 97.3%, and 96.8%, respectively. However, the included studies of the meta-analysis mainly resulted from the target population who had already been screen on US, and the MRI was used as an adjunct tool which probably contributed to a risk of selection bias. Therefore, the predictive accuracy of MRI in defining the grade of PAS might not be clearly demonstrated by these results and further more evidence is needed.

There have been more studies that assess and compare the predictive performance of US and MRI in PAS. D’Antonio et al. in a recent meta-analysis reported that the pooled sensitivity and specificity of US in diagnosing PAS were 90.7% and 96.9% [29], and subsequently reported that the pooled sensitivity and specificity of MRI in diagnosing PAS were 94.4% and 84% in another study [30]. One study comparing the diagnostic value of these two imaging tools within in one systematic review was published in 2013 by Meng et al. [31], reported that pooled sensitivity of diagnostic value for US and MRI was 83% and 82%, respectively, corresponding pooled specificity for US and MRI was 95% and 88%, respectively. Although these reviews conclude that both US and MRI were equally effective for identification of PAS, evaluation of the diagnostic performance of US and MRI were conducted on separate patient populations, which unavoidably represents a source of bias.

To precisely evaluate the predictive accuracy for PAS identification between US and MRI, we only included studies in which US and MRI were conducted within the identical patient population. In this present study, we found that the pooled sensitivity and specificity for US was 90% and 83% and that for MRI was 89% and 87%, respectively. We concluded that both of the two image modalities (US and MRI) have an acceptable diagnostic value with high sensitivity and specificity for the prenatal identification of PAS, which is in accordance with the previous reports [31]. The pooled diagnostic sensitivity (p = 0.808) and specificity (p = 0.413) between US and MRI were not significantly different in the detection of PAS. Further analysis and calculation based on SROC curves showed that AUC of US and MRI was 0.935 and 0.932, and the Q* indices were 0.872 and 0.867 for US and MRI, respectively. Z value test was then used to compare the significance of the two imaging techniques for diagnosing PAS by analyzing the AUC of SROC curve, and the results revealed the Z value of 0.1299 with a corresponding p value of 0.5517, indicating that there was no statistical difference for US and MRI in predictive accuracy for PAS. To further evaluate the heterogeneity of results, we conducted subgroup analyses (i.e., retrospective vs prospective, early vs recent study) and results also showed that there was no statistical significance in sensitivity, specificity as well as AUC of SROC curves between US and MRI. All results revealed that both US and MRI possessed equal accuracy in the prenatal diagnosis of PAS, and one modality has not demonstrated superiority over the other.

Although some studies suggest that MRI may be more useful than US and expert opinion and guidelines [19, 32] recommend the use of MRI in certain scenarios (i.e., posterior placental implantation, inconclusive US assessment), more evidence is required for validation. Dwyer et al. [11] found there was no significant difference in diagnostic accuracy for prenatal identification of PAS between US and MRI, and also the performance of the two imaging modalities was not impaired by the placental site (i.e., posterior placentas). Additionally, one recent study performed by Einerson et al. [33] to evaluate the diagnostic value of MRI in the identification of PAS as a secondary tool revealed that MRI only results in 19% of correct, clinically meaningful change in the US-based PAS diagnosis, and yet an incorrect change of 17% of US-based diagnosis and a 21% of incorrect confirmation of US-based diagnosis. Furthermore, MRI offered no identifiable benefit for cases that the placenta was located in posterior or lateral position. Therefore, some researchers advocated that the employment of MRI as a secondary tool to US in prenatal identification of PAS should not be routinely recommended, and obstetric management should not be altered based on MRI findings if there is a strong indication of PAS by US with several positive features with good PPV [16, 34]. Further definite studies are required to demonstrate the diagnostic accuracy of MRI for more complete evidence.

The wide difference in diagnostic accuracy between studies can be attributed to multiple factors, including small sample size, type of study, variability of study inclusion and exclusion criteria, reference ultrasound or MRI signs of PAS, operator’s experience, equipment, gestational age and confirmative diagnosis by the gold standard modality of histopathology which is often unavailable in adherent placenta or conservatively managed cases. In particular, there has been no consensus on the reference findings on US or MRI that are indicative of PAS, and the predictive performance of these two imaging modalities is largely dependent on the experience and expertise of the sonographers and radiologists. This is further limited by the absence of a standardized training program in imaging interpretation and the relative rarity of the PAS disease.

Better sensitivity but a comparably lower specificity of either US or MRI techniques for the diagnosis of PAS, perhaps because once one feature of the diagnostic criteria was present, the diagnosis of PAS was considered, consequently the number of false positives was increased, and specificity of the test reduced. The most recent retrospective cohort study [35] exhibited that a comprehensive scoring system by combing US, MRI and clinical risk factors was superior to either modality when used alone, suggesting that a complementary approach by combing several characteristic markers from various clinical parameters into a structured and quantified system might improve the diagnostic accuracy.

There are several limitations of this study. First, we acknowledge that including the assessment of topography and grade of invasive placentas can better qualify and quantify the diagnostic efficacy of US and MRI. However, this attempt is limited by the scarcity of such information in included studies. Second, the accuracy of the meta-analysis is largely dependent on the reliability of the included research; although subgroup analysis has been performed to elucidate the heterogeneity of results, this remains as a source of unavoidable bias. Third, we did not take into account some factors such as maternal characteristics (i.e., inclusion criterion of the different studies, placental location, BMI) and gestational age for imaging that may impact the predictive performance of these two imaging modalities.

The strengths of our study include the following: (1) The comparison of diagnostic accuracy between US and MRI for the identification of PAS was examined in the identical patient population for the first time. (2) We included a larger population compared to previous publications to assess the predictive performance of US and MRI. (3) Majority of the included studies showed that depictions of both US and MRI were evaluated by a skilled and experienced sonographer or radiologist without knowledge of the pathologic results or surgical findings. (4) Both sensitivity analysis and subgroup analysis revealed the similar result that there was no statistical significance of diagnostic accuracy between US and MRI, indicating the robust outcome in the present study.

In conclusion, both US and MRI demonstrate equal efficacy in diagnosing PAS and one modality has not demonstrated superiority over the other. However, routine employment of MRI with relatively high expense in the prenatal identification of PAS should not be recommended, and further definite studies are required to demonstrate the diagnostic accuracy of MRI for more complete evidence. The development of standardized training programs and comprehensive scoring system using US, MRI and clinical risk factors can help to improve accurate prenatal diagnosis of PAS.

Availability of data and materials

All data generated or analyzed during this study are included in this published article.

Abbreviations

AUC:

Area under the SROC curve

DOR:

Diagnostic odds ratio

LR−:

Negative likelihood ratios

LR+:

Positive likelihood ratios

MRI:

Magnetic resonance imaging

PAS:

Placenta accrete spectrum

QUADAS:

Quality Assessment of Diagnostic Accuracy Studies

SROC:

Summary receiver operating characteristic

US:

Ultrasonography

References

  1. Solheim K, Esakoff T, Little S, Cheng Y, Sparks T, Caughey A (2011) The effect of cesarean delivery rates on the future incidence of placenta previa, placenta accreta, and maternal mortality. J Matern Fetal Neonatal Med 24:1341–1346

    Article  Google Scholar 

  2. Derman A, Nikac V, Haberman S, Zelenko N, Opsha O, Flyer M (2011) MRI of placenta accreta: a new imaging perspective. AJR Am J Roentgenol 197:1514–1521

    Article  Google Scholar 

  3. Creanga A, Bateman B, Butwick A et al (2015) Morbidity associated with cesarean delivery in the United States: Is placenta accreta an increasingly important contributor? Am J Obstet Gynecol 213:384.e381–311

    Article  Google Scholar 

  4. Zelop C, Harlow B, Frigoletto F, Safon L, Saltzman D (1993) Emergency peripartum hysterectomy. Am J Obstet Gynecol 168:1443–1448

    Article  CAS  Google Scholar 

  5. O’Brien J, Barton J, Donaldson E (1996) The management of placenta percreta: conservative and operative strategies. Am J Obstet Gynecol 175:1632–1638

    Article  CAS  Google Scholar 

  6. Jauniaux E, Bhide A, Kennedy A, Woodward P, Hubinont C, Collins S (2018) FIGO consensus guidelines on placenta accreta spectrum disorders: Prenatal diagnosis and screening. Int J Gynaecol Obstet 140:274–280

    Article  Google Scholar 

  7. Bailit J, Grobman W, Rice M et al (2015) Morbidly adherent placenta treatments and outcomes. Obstet Gynecol 125:683–689

    Article  Google Scholar 

  8. Zamora J, Abraira V, Muriel A, Khan K, Coomarasamy A (2006) Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med Res Methodol 6:31

    Article  Google Scholar 

  9. Levine D, Hulka C, Ludmir J, Li W, Edelman R (1997) Placenta accreta: evaluation with color Doppler US, power Doppler US, and MR imaging. Radiology 205:773–776

    Article  CAS  Google Scholar 

  10. Masselli G, Brunelli R, Casciani E et al (2008) Magnetic resonance imaging in the evaluation of placental adhesive disorders: correlation with color Doppler ultrasound. Eur Radiol 18:1292–1299

    Article  Google Scholar 

  11. Dwyer B, Belogolovkin V, Tran L et al (2008) Prenatal diagnosis of placenta accreta: sonography or magnetic resonance imaging? J Ultrasound Med 27:1275–1281

    Article  Google Scholar 

  12. Mausour SM, Elkhyat WM (2011) Placenta Previa-accreta: Do we need MR imaging? Egypt J Radiol Nuclear Med 42:433–442

    Article  Google Scholar 

  13. Lim P, Greenberg M, Edelson M, Bell K, Edmonds P, Mackey A (2011) Utility of ultrasound and MRI in prenatal diagnosis of placenta accreta: a pilot study. AJR Am J Roentgenol 197:1506–1513

    Article  Google Scholar 

  14. Elhawary T, Dabees N, Youssef M (2013) Diagnostic value of ultrasonography and magnetic resonance imaging in pregnant women at risk for placenta accreta. J Matern Fetal Neonatal Med 26:1443–1449

    Article  Google Scholar 

  15. Peker N, Turan V, Ergenoglu M et al (2013) Assessment of total placenta previa by magnetic resonance imaging and ultrasonography to detect placenta accreta and its variants. Ginekol Pol 84:186–192

    Article  Google Scholar 

  16. Riteau A, Tassin M, Chambon G et al (2014) Accuracy of ultrasonography and magnetic resonance imaging in the diagnosis of placenta accreta. PLoS One 9:e94866

    Article  Google Scholar 

  17. Algebally A, Yousef R, Badr S, Al Obeidly A, Szmigielski W, Al Ibrahim A (2014) The value of ultrasound and magnetic resonance imaging in diagnostics and prediction of morbidity in cases of placenta previa with abnormal placentation. Pol J Radiol 79:409–416

    Article  Google Scholar 

  18. Satija B, Kumar S, Wadhwa L et al (2015) Utility of ultrasound and magnetic resonance imaging in prenatal diagnosis of placenta accreta: a prospective study. Indian J Radiol Imaging 25:464–470

    Article  Google Scholar 

  19. Balcacer P, Pahade J, Spektor M, Staib L, Copel J, McCarthy S (2016) Magnetic resonance imaging and sonography in the diagnosis of placental invasion. J Ultrasound Med Off J Am Inst Ultrasound Med 35:1445–1456

    Google Scholar 

  20. Rezk M, Shawky M (2016) Grey-scale and colour Doppler ultrasound versus magnetic resonance imaging for the prenatal diagnosis of placenta accreta. J Matern Fetal Neonatal Med 29:218–223

    Article  Google Scholar 

  21. Daney de Marcillac F, Molière S, Pinton A et al (2016) Accuracy of placenta accreta prenatal diagnosis by ultrasound and MRI in a high-risk population. J Gynecol Obstet Biol Reprod (Paris) 45:198–206

    Article  CAS  Google Scholar 

  22. Budorick N, Figueroa R, Vizcarra M, Shin J (2017) Another look at ultrasound and magnetic resonance imaging for diagnosis of placenta accreta. J Matern Fetal Neonatal Med 30:2422–2427

    Article  CAS  Google Scholar 

  23. Kumar I, Verma A, Ojha R, Shukla R, Jain M, Srivastava A (2017) Invasive placental disorders: a prospective US and MRI comparative analysis. Acta Radiol 58:121–128

    Article  Google Scholar 

  24. Ayati S, Leila L, Pezeshkirad M et al (2017) Accuracy of color Doppler ultrasonography and magnetic resonance imaging in diagnosis of placenta accreta: a survey of 82 cases. Int J Reprod Biomed 15:225–230

    PubMed  PubMed Central  Google Scholar 

  25. Romeo V, Sarno L, Volpe A et al (2019) US and MR imaging findings to detect placental adhesion spectrum (PAS) in patients with placenta previa: a comparative systematic study. Abdom Radiol (NY) 44:3398–3407

    Article  CAS  Google Scholar 

  26. Xia H, Ke S, Qian R, Lin J, Li Y, Zhang X (2020) Comparison between abdominal ultrasound and nuclear magnetic resonance imaging detection of placenta accreta in the second and third trimester of pregnancy. Medicine (Baltimore) 99:e17908

    Article  Google Scholar 

  27. Tabsh K, Brinkman C, King W (1982) Ultrasound diagnosis of placenta increta. J Clin Ultrasound 10:288–290

    Article  CAS  Google Scholar 

  28. Familiari A, Liberati M, Lim P et al (2018) Diagnostic accuracy of magnetic resonance imaging in detecting the severity of abnormal invasive placenta: a systematic review and meta-analysis. Acta Obstet Gynecol Scand 97:507–520

    Article  Google Scholar 

  29. D’Antonio F, Iacovella C, Bhide A (2013) Prenatal identification of invasive placentation using ultrasound: systematic review and meta-analysis. Ultrasound Obstet Gynecol 42:509–517

    Article  CAS  Google Scholar 

  30. D’Antonio F, Iacovella C, Palacios-Jaraquemada J, Bruno C, Manzoli L, Bhide A (2014) Prenatal identification of invasive placentation using magnetic resonance imaging: systematic review and meta-analysis. Ultrasound Obstet Gynecol Off J Int Soc Ultrasound Obstet Gynecol 44:8–16

    Article  CAS  Google Scholar 

  31. Meng X, Xie L, Song W (2013) Comparing the diagnostic value of ultrasound and magnetic resonance imaging for placenta accreta: a systematic review and meta-analysis. Ultrasound Med Biol 39:1958–1965

    Article  Google Scholar 

  32. Berkley E, Abuhamad A (2013) Prenatal diagnosis of placenta accreta: Is sonography all we need? J Ultrasound Med Off J Am Inst Ultrasound Med 32:1345–1350

    Google Scholar 

  33. Einerson B, Rodriguez C, Kennedy A, Woodward P, Donnelly M, Silver R (2018) Magnetic resonance imaging is often misleading when used as an adjunct to ultrasound in the management of placenta accreta spectrum disorders. Am J Obstet Gynecol 218:618.e611-618.e617

    Article  Google Scholar 

  34. McLean L, Heilbrun M, Eller A, Kennedy A, Woodward P (2011) Assessing the role of magnetic resonance imaging in the management of gravid patients at risk for placenta accreta. Acad Radiol 18:1175–1180

    Article  Google Scholar 

  35. Knight J, Lehnert S, Shanks A et al (2018) A comprehensive severity score for the morbidly adherent placenta: combining ultrasound and magnetic resonance imaging. Pediatr Radiol 48:1945–1954

    Article  Google Scholar 

Download references

Acknowledgements

This study was supported by a grant from Natural Science Foundation of Science and Technology Commission of Shanghai Municipality (No. 22ZR1438700), China.

Funding

This study was supported by a grant from Natural Science Foundation of Science and Technology Commission of Shanghai Municipality (No. 22ZR1438700), China.

Author information

Authors and Affiliations

Authors

Contributions

SH has significant statistical expertise. The scientific guarantor of this publication is NZ. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Yu Zhang or Ning Zhang.

Ethics declarations

Ethics approval and consent to participate

Institutional Review Board approval was not required because this study belongs to meta-analysis. Written informed consent was not required for this study because this study belongs to meta-analysis; the data are from various published literatures.

Consent for publication

Not applicable.

Competing interests

The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hong, S., Le, Y., Lio, K.U. et al. Performance comparison of ultrasonography and magnetic resonance imaging in their diagnostic accuracy of placenta accreta spectrum disorders: a systematic review and meta-analysis. Insights Imaging 13, 50 (2022). https://doi.org/10.1186/s13244-022-01192-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13244-022-01192-w

Keywords