Skip to main content

Table 2 MRI protocol for the study of bone sarcomas in a 3-T scanner

From: Multiparametric MRI evaluation of bone sarcomas in children

 

Slice number

FOV (mm)

Voxel size (mm)

Slice thickness (mm)

Gap

TR (ms)

TE (ms)

NEX

Flip angle

TA (m:s)

b value

Coronal T2 DIXON

35

429

0.6 × 0.6

3

0

3300

80

1

90

2:34

 

Coronal T1 TSE

35

429

0.5 × 0.6

3

0

700

10

1

90

3:03

 

Sagittal T2 DIXON

52

300

0.6 × 0.7

1.5

0

3200

60

1

90

2:38

 

Axial T2 DIXON

60

120

0.6 × 0.7

3

0

3600

60

1

90

2:48

 

Axial T1 TSE

60

120

0.5 × 0.6

3

0

700

10

1

90

2:09

 

Axial DWI

60

120

1.1 × 1.1

3

0

9000

78

1

90

5:08

0/500/800

THRIVE T1 TFE

75 (30DYN)

300

1.5 × 1.5

1.5

0

4

2

1

10

5:08

 

Sagittal T1 FS + C

52

300

0.8 × 0.8

1.5

0

600

10

1

90

2:19

 

Axial T1 FS + C

60

120

0.5 × 0.6

3

0

700

10

1

90

3:28

 
  1. 30DYN = 30 series after contrast administration. C = contrast agent. DWI = diffusion-weighted imaging. FFE = fast field echo. FOV = field of view. FS = fat suppression. m:s = minutes:seconds. ms = milliseconds. NEX = number of excitations. TA = acquisition time. TE = echo time. TR = repetition time. TSE = turbo spin echo