The assessment of image quality and diagnostic value in X-ray images: a survey on radiographers’ reasons for rejecting images

Abstract

Background

Assessing the quality of diagnostic images is subjective and influenced by factors such education, skills, and experience of the assessor. This study aims to explore the radiographers’ assessments of medical usefulness or rejection of X-ray images in specific cases.

Results

Eighty-one radiographers from different countries responded to the questionnaire distributed online at the EFRS research HUB at ECR 2020 (a 15% response rate). Forty-two percent of the respondents practiced in the UK and Ireland. In addition to rejecting or keeping images in the presented 30 cases and giving a main reason for the images rejected, the participants explained their choice using comments, 1176 comments were obtained. Sixty percent of the comments were on kept images. The respondents kept on average 63% of the images. In the “Keep”, “Could keep”, and “Reject” categories on average 84%, 63% and 43% of images were kept respectively. The most common reasons given for rejecting an image were suboptimal positioning and centering. Potential diagnostic value and radiation protection were indicated as reasons to keep an image perceived as of low quality reported in n = 353 and n = 33 comments respectively.

Conclusions

There is an agreement internationally on what makes a good quality X-ray image. However, the opinion on medical usefulness of images of low or poor quality compared to image criteria varies. Diagnostic capability and radiation protection was the rationale used for keeping images not fulfilling image criteria. There seems to be a need for diagnostic quality to be included in image assessment in clinical practice.

Key points

  • Radiographers internationally highly agree on what makes a good X-ray image.

  • Assessing images mainly against image criteria may lead to more rejections.

  • Training for radiographers in image quality assessment and regular reject analysis might reduce retake rate.

  • Dialogue between radiologists and radiographers on diagnostic image quality is important in clinical practice.

Introduction

The quality of a radiographic image influences diagnostic accuracy and subsequent clinical management of the patient [1]. A radiographic image is accepted as good quality when certain technical qualities are satisfied, and the image considered of diagnostic value [2]. An image of low quality has the possibility of rejection and the radiological procedure repeated in attempt to produce imaging of diagnostic value [3]. Rejects and subsequent retakes of radiographic images expose both patients and radiological personnel to unnecessary ionising radiation [4]. Over exposure to ionising radiation increases risk of probability of radiation induced stochastic effects [5]. Repeated imaging further results to higher usage of both human and radiological resources [6]. Jones et al. [3] state that rejected images are integral to X-ray imaging, where patient positioning and alignment are vital components of image quality. During a radiological procedure, patient motion, positioning, and artefacts unique to the image receptor technology can result in repeated images [3] and are reported as the major reasons for image retakes [7]. Examining the underlying causes for image rejections assists to identify any technical or training issues required in a radiology department [8]. This further align with the As Low As Reasonably Achievable, (ALARA) principle and global initiatives of imaging wisely for proper use of radiological resources [3, 9].

Image quality assessment is reported to be subjective and significantly differs between and among radiographers and radiologists. Dunn and Rogers [10] found that radiologists retained as useful 50% more of the images that showed positioning errors compared to radiographers, indicating a difference in clinical judgement between the two groups. Kjelle et al. [11] further report that although there is a professional agreement on the assessment of good quality images, radiologists tend to accept a higher number of images that radiographers reject and considered to be of low quality. Mount [12] suggests that radiographers generally tend to favour technical characteristics of the image. Radiographers are inclined to assess the quality of imaging for proper positioning, adequate exposure, patient motion blur and other defects that could potentially affect diagnosis [13]. Artefacts and processing errors are also reported for reasons radiographers reject imaging [7, 8, 14]. Technical criteria in image quality are however hard to fulfil and could lead to radiographers’ premature rejection of images that potentially provide radiologists with information of diagnostic value [12].

Several factors are reported to influence rejects of imaging among radiographers including skills and the inter-subjectivity in assessment, and attitude towards, both technical and clinical image quality criteria [12, 15]. Waaler and Hofmann [15] highlight inter-subjectivity of radiographers’ perception of image quality to be a challenge in the effort to reduce reject rate. Having higher years of work experience is reported to have some influence on how radiographers assess images, with more experienced radiographers assessing more than merely radiographic imaging criteria [16]. The difference in image quality assessment due to years of experience is also shown among radiologists. Saade et al. [17] reports of senior radiologists scoring image quality and image technique as acceptable 32% and 27% respectively more than junior radiologists. The education of the radiographers within a particular radiography subspecialty also influences assessment of image quality. Mercieca, et al. [18] reports of radiographers with no specific mammography qualifications rejecting more images than those with a specialisation.

This study is part of a larger project investigating different aspects of image quality assessments in clinical practice that may influence retake rate. The aim of this study was to survey among radiographers in medical imaging internationally to explore assessments of usability of X-ray images and the need for image retake in specific cases and compare them with results from Norway.

Methods

A semi-structured questionnaire was distributed through the European Federation of Radiographer Societies (EFRS) research hub at the European congress of radiology (ECR) in 2020. All radiographers attending the online congress were invited to participate. Five hundred and forty-six radiographers visited the research hub in July 2020. The respondents gave written, informed consent.

The questionnaire included 30 clinical cases (5 paediatric) in addition to questions collecting demographics of the respondents. Cases were provided from the picture and archiving system of one hospital in Norway. Each case included one image and a short referral text. The questionnaire included three chest, five spine, fourteen upper-limb, and eight lower-limb X-ray cases, representing examinations shown to have high rejection rates [19]. The images were pre-assessed into three categories (Keep, could keep and reject) by the authors based on European guidelines on quality criteria for diagnostic radiographic images [1]. In the categories reject and could accept cases the images flaws were: suboptimal positioning (36.4%; one case of wrong extremity imaged), collimation (22.7%), exposure error (18.2%), centering (13.6%) and artifacts (9.1%).

In each case, the respondents were to choose to either keep or reject the image. When rejecting an image, the respondents were to choose from a predefined list with the following options on the main flaw of the image: positioning is suboptimal, centering is suboptimal, artefact in the image, collimation is suboptimal, or exposure error as the main reason for rejection. In addition, there was an option to comment on why the image was kept or rejected, thus providing both qualitative and quantitative data.

Analysis

R (R Core Team, 2020) was used to perform 2-tailed chi-squared tests to explore group differences.

The written comments were analysed grouping comments on kept and rejected images and tabulating comments into the following categories. Fulfilment of image criteria, diagnostic value, radiation protection, discuss with colleague/radiologist/referrer, and other.

The Norwegian Centre for Research Data approved this study for data handling (Reference: 987929).

Results

In total, 81 radiographers from 21 different countries responded to the survey, a response rate of 15% of participants in the EFRS research HUB. The majority of respondents were from European countries (76%), among these most respondents were working in Ireland and the UK. Demographics of the respondents are presented in Table 1.

Table 1 Demographics of respondents in the survey

All participants answered all cases and documented 1176 comments. The respondents kept on average 63% of the images, in the “Keep” category on average 84% (SD 10) of the images were kept. In the “Could keep” category on average 63% (SD 22) of the images were kept, and in the “Reject” category the average keep rate was 43% (SD 23). There were no significant difference in rate of kept images comparing years of experience, or when comparing respondents from European and non-European countries. The Chi-squared test was used to compare European radiographers in this study to radiographers and radiologist in a previous identical survey conducted in Norway [11] (Table 2). Radiographers from the UK and Ireland kept significantly more images than those from Norway in the “Could keep” and “Reject” categories. While radiographers from the rest of Europe kept significantly more images than those from Norway in all categories. There was no significant difference between radiographers in Europe and the radiologist in the survey conducted in Norway. There was no significant difference between radiographers in UK/Ireland and the rest of Europe (data not shown).

Table 2 Comparing the keep rate for radiographers in Europe to radiographers and radiologists from Norway using the two-tailed chi-square test with 1 degree of freedom. Data from Norway is previously published [11]

Reasons for rejecting an image

Figure 1 gives an overview of the reasons given for rejecting images in all categories. Positioning is suboptimal (47%) and centering is suboptimal (22%) were the main reasons given for rejecting an image, which is a higher rate than the initial assessment where these reasons were used in 27% and 10% of images respectively.

Fig. 1
figure 1

Overview of main reasons for rejection of images

Comments on choice to keep or reject

Seven hundred and eleven comments were documented on kept images, while 456 comments were on rejected images; all images had comments (2–49 per case). Comments were sorted into two categories depending on whether the elaboration was written for a kept or rejected image. Table 3 presents the sub-categories in each main category.

Table 3 Overview of categories of comments on kept and rejected images

For kept images, the most common comments (n = 353) were on how the image could have diagnostic value even though the image criteria were not fulfilled. In other cases, the elaboration was used to comment that image criteria were met (n = 292). In addition, the thirty-three comments discussed that radiation protection would be a reason to keep an image of low quality and five comments referred to confer with a radiologist or referring physician before deciding to keep or reject.

For rejected images, most comments were on lack of image criteria fulfilment (n = 439), including missing markers. Twelve comments discussed lack of diagnostic value, while four elaborations were on the possibility to keep the image if the other images in the series compensated for the information missing in the current image.

Discussion

In this study, we found that radiographers kept on average 63% of the cases. Images pre-assessed as of low quality were rejected most often. The comments on rejected images focused on lack of image criteria fulfilment or low diagnostic quality. While comments on kept images focused on diagnostic value despite image criteria not being fulfilled or stating that image quality criteria was fulfilled in images of good quality.

Comparing radiographers in different parts of Europe to radiographers and radiologists in Norway showed that radiographers from Norway rejected a significantly higher proportion of images. There was no significant difference between radiologists from Norway and radiographers from other European countries.

Image criteria verses diagnostic quality

The reported reasons for a rejected image in our study were attributed to technical skills or human errors. Insufficient patient positioning (47%) was the main given reason for rejection. The image criteria specify important anatomical structures that should be visible on a radiograph to aid accurate diagnosis which of these criteria fundamentally depend on accurate patient positioning [2]. Therefore, proper positioning of the patient is one important required factor in production of high-quality images of diagnostic value. Our study does however indicate from the comments that the radiographers would also consider diagnostic value in assessing whether to keep or reject an image. The radiographers’ role is to ensure that imaging of diagnostic quality is obtain and the radiologists interprets the images and provide diagnosis. Where a low quality image provides sufficient information for accurate diagnosis, the radiologist is able to answer the clinical question and thus would be unnecessary to repeat imaging [17]. However, images rejected by radiographers are mostly not reviewed by the radiologist [19]. Even though diagnostic value would mainly be the radiologist domain radiographers need to consider the images diagnostic value if an image does not fully fulfil image criteria, sometimes in dialog with the radiologists and in other cases with the support of another radiographer. The difference in rejects of images among radiographers and radiologist regarding the issue of technical versus diagnostic capability is reported. Mount [12] reports that radiographers’ focus on technical characteristics of an image can lead to premature rejection of images of diagnostic value. Evaluation on technical criteria can also be difficult to satisfy thus leading to more rejects [13]. To prevent unnecessary imaging retakes it is important that discussions to weigh technical characteristics verse diagnostic benefits between radiographers and radiologists occur [19]. Thus, training in image quality assessment is central for radiographers.

Difference in training

Our study shows variation in rejection rates among radiographers in the included European countries. Many factors such as training or work experience as reported by several studies could have an influence [16,17,18]. Training of radiographers to improve the technical skills and operation of equipment is vital to reduce imaging errors [20]. McFadden et al. [21] advocated for standardisation of education and training within Europe including protocols and exposure parameters to ensure that there is continued adherence to the ALARA principle. Practical aspects in routine clinical practice are also vital to reduce technical errors. Lin et al. [7], recommends introduction of position and equipment technical guidelines on required patient positioning and adding poster reminders to radiographers to assist to reduce positioning errors and rejects. Current advancement in deep learning and artificial intelligence should possibly resolve or reduce rejects caused by technical or human errors particularly patient position when implemented [22].

Radiation protection

Our study shows from the comments that radiation protections was also an important factor for keeping an image of lower quality. Rejected images increase radiation exposure to patients, increasing the risk of stochastic effects [4]. The risk of stochastic effects are dependent on age, sex and individual health status [5]. The accumulative dose exposure from previous radiological procedures should also be considered before rejecting and repeating the radiographic image [23]. This further highlights the importance of discussions with radiologists particularly when assessing the need for retakes in the vulnerable groups such as in paediatric imaging. Ensuring patient radiation safety and minimising exposure of ionising radiation or other harm in imaging is fundamental to healthcare in radiology [4].

Limitations of study

The response rate was low only 15%. A low response rate reduces strength of analysis with regards to producing a representative result [24]. Due to the Covid-19 pandemic, the ECR 2020 was held virtually, and the distributed survey was displayed to visitors at an online research hub along with several other surveys. The response rate was calculated from the number of participants who visited the hub web page. There may have been visitors to the page who did not participate in any surveys and thus the rate may have been under reported. It was also not possible to prompt individuals directly or send reminders to visitors of the page. This may have influenced the number of potential participants reached, further contributing to the low response rate. Additionally, the initial assessment of image quality of the cases in the survey is based on guidelines and subjective assessment of the authors. This approach has limitations. Discussing the cases with radiologists and other radiographers could have helped to strengthen validity. The survey was presented in English only. This may have hindered some radiographers’ participation causing skewedness in number of participants from English speaking countries in Europe. Language could have further caused misunderstanding of the presented cases for non-native English-speaking participants. On the other hand, English is the language of official communication and correspondence at the ECR. The participants would thus be expected to understand English.

Conclusion

This study shows that there is an agreement internationally on what makes a good quality X-ray image. However, the opinion on medical usefulness of images of low or poor quality compared to image criteria varies. The radiographers considered the lack of fulfilment of image criteria as a reason for rejecting an image whereas diagnostic capability was an important reason for keeping images of low quality. There seems to be a difference in image rejection rate in this survey based on whether radiographers strictly adhere to the image criteria rather than assessing diagnostic quality. Our study suggests that radiographers from Norway reject more images based on lack of fulfilment of image criteria than their international colleagues. Radiologists in Norway seems be on a similar level of rejection as radiographers internationally. In addition to regular image reject analysis and training for radiographers in image assessment, we highly recommend radiographers and radiologists to communicate on image quality in clinical practice, and for diagnostic quality to be included in image criteria assessment.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ALARA:

As Low As Reasonably Achievable

ECR:

European congress of radiology

EFRS:

European Federation of Radiographer Societies

UK:

United Kingdom

References

  1. Carmichael JHE, Maccia C, Moores BM et al (2000) European guidelines on quality criteria for diagnostic radiographic images. EU Publication

    Google Scholar 

  2. Maccia C, Moores BM, Wall BF (1997) The 1991 CEC trial on quality criteria for diagnostic radiographic images: detailed results and findings. EN Luxembourg Office for Official Publication of the European Communities

    Google Scholar 

  3. Jones AK, Heintz P, Geiser W et al (2015) Ongoing quality control in digital radiography: report of AAPM Imaging Physics Committee Task Group 151. Med Phys 42:6658–6670. https://doi.org/10.1118/1.4932623

    Article  PubMed  Google Scholar 

  4. International Atomic Energy Agency (IAEA) (2018) Radiation protection and safety in medical uses of ionizing radiation, IAEA Safety Standards Series No. SSG-46. IAEA: Vienna. https://www-pub.iaea.org/MTCD/Publications/PDF/PUB1775_web.pdf. Accessed 10 Aug 2021

  5. International Commission on Radiological Protection ICRP (2013) Radiological protection in paediatric diagnostic and interventional radiology publication 121. Ann ICRP 42:1–63. https://doi.org/10.1016/j.icrp.2012.10.001

    Article  Google Scholar 

  6. Almalki AA, Manaf RA, Juni MH, Hayati KS, Noor NM, Gabbad AAM (2017) A Systematic review on repetition rate of routine digital radiography. Int J Curr Res 9:46325–46330

    Google Scholar 

  7. Lin C, Chan P, Huang K, Lu C, Chen Y, Lin Chen Y (2016) Guidelines for reducing image retakes of general digital radiography. Adv Mech Eng. https://doi.org/10.1177/1687814016644127

    Article  Google Scholar 

  8. Taylor N (2015) The art of rejection: comparative analysis between Computed Radiography (CR) and Digital Radiography (DR) workstations in the Accident & Emergency and General radiology departments at a district general hospital using customised and standardised reject criteria over a three year period. Radiography 21:236–241. https://doi.org/10.1016/j.radi.2014.12.003

    Article  Google Scholar 

  9. Brink JA, Amis SE (2010) Image wisely: a campaign to increase awareness about adult radiation protection. Radiology 257:601–602. https://doi.org/10.1148/radiol.10101335

    Article  PubMed  Google Scholar 

  10. Dunn MA, Rogers AT (1998) X-ray film reject analysis as a quality indicator. Radiography 4:29–31. https://doi.org/10.1016/S1078-8174(98)80027-8

    Article  Google Scholar 

  11. Kjelle E, Schanche AK, Hafskjold L (2021) To keep or reject, that is the question—a survey on radiologists and radiographers’ assessments of plain radiography images. Radiography (Lond) 27:115–119. https://doi.org/10.1016/j.radi.2020.06.020

    Article  CAS  Google Scholar 

  12. Mount J (2016) Reject analysis: A comparison of radiographer and radiologist perceptions of image quality. Radiography 22:e112–e117. https://doi.org/10.1016/j.radi.2015.12.001

    Article  Google Scholar 

  13. Whaley JS, Pressman BD, Wilson JR, Bravo L, Sehnert WJ, Foos DH (2013) Investigation of the variability in the assessment of digital chest X-ray image quality. J Digit Imaging 26:217–226. https://doi.org/10.1007/s10278-012-9515-1

    Article  PubMed  Google Scholar 

  14. Sl L, Mak AS, Lam W, Chau C, Lau K (2004) Reject analysis: a comparison of conventional film–screen radiography and computed radiography with PACS. Radiography 10:183–187. https://doi.org/10.1016/j.radi.2004.03.014

    Article  Google Scholar 

  15. Waaler D, Hofmann B (2010) Image rejects/retakes-radiographic challenges. Radiat Prot Dosimetry 139:375–379. https://doi.org/10.1093/rpd/ncq032

    Article  CAS  PubMed  Google Scholar 

  16. Waaler D, Hammer S, Langdalen C, Haug LTH (2017) How radiographers visually perceive X-ray images with the task of accepting or rejecting them—a pilot study. Radiography Open 3(1):10. https://doi.org/10.7577/radopen.1997

    Article  Google Scholar 

  17. Saade C, Siblini L, Karout L et al (2021) To repeat or not to repeat: Radiologists demonstrated more decisiveness than their fellow radiographers in reducing the repeat rate during mobile chest radiography. Radiography (Lond) 27:304–309. https://doi.org/10.1016/j.radi.2020.09.003

    Article  CAS  Google Scholar 

  18. Mercieca N, Portelli JL, Jadva-Patel H (2017) Mammographic image reject rate analysis and cause—a National Maltese Study. Radiography (Lond) 23:25–31. https://doi.org/10.1016/j.radi.2016.07.004

    Article  CAS  Google Scholar 

  19. Atkinson S, Neep M, Starkey D (2020) Reject rate analysis in digital radiography: an Australian emergency imaging department case study. J Med Radiat Sci 67:72–79. https://doi.org/10.1002/jmrs.343

    Article  PubMed  Google Scholar 

  20. Hofmann B, Rosanowsky TB, Jensen C, Wah KH (2015) Image rejects in general direct digital radiography. Acta Radiol Open 4:2058460115604339. https://doi.org/10.1177/2058460115604339

    Article  PubMed  PubMed Central  Google Scholar 

  21. Mc Fadden S, Roding T, de Vries G, Benwell M, Bijwaard H, Scheurleer J (2018) Digital imaging and radiographic practise in diagnostic radiography: an overview of current knowledge and practice in Europe. Radiography (Lond) 24:137–141. https://doi.org/10.1016/j.radi.2017.11.004

    Article  CAS  Google Scholar 

  22. Ohta Y, Matsuzawa H, Yamamoto K, Enchi Y, Kobayashi T, Ishida T (2021) Development of retake support system for lateral knee radiographs by using deep convolutional neural network. Radiography (Lond). https://doi.org/10.1016/j.radi.2021.05.002

    Article  Google Scholar 

  23. Frush D (2021) The cumulative radiation dose paradigm in pediatric imaging. Br J Radiol 94(1126):20210478. https://doi.org/10.1259/bjr.20210478

    Article  PubMed  Google Scholar 

  24. Altman DG (2020) Practical statistics for medical research, 2nd edn. Chapman and Hall, London

    Google Scholar 

Download references

Acknowledgements

We would like to thank Linda Hafskjold and Ann Kristin Schanche for their contribution in the planning of this study and for their help in the development of the questionnaire. We would like to thank the EFRS for including our survey at the online research hub at European Congress of Radiology (ECR) 2020. Further, we would like to thank the participants for taking time to visit the online EFRS research hub at ECR 2020 and responding to our survey.

Funding

No funding received.

Author information

Authors and Affiliations

Authors

Contributions

First author- Planning the study and developing the questionnaire, distributing the questionnaire online, analysis, writing the manuscript. Last author- Developing the questionnaire, analysis, writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Elin Kjelle.

Ethics declarations

Ethics approval and consent to participate

The Norwegian Centre for Research Data approved this study for data handling (Reference: 987929). Participants gave written consent to participate before starting the survey.

Consent for publication

Participants gave written consent for publication before starting the questionnaire.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kjelle, E., Chilanga, C. The assessment of image quality and diagnostic value in X-ray images: a survey on radiographers’ reasons for rejecting images. Insights Imaging 13, 36 (2022). https://doi.org/10.1186/s13244-022-01169-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13244-022-01169-9

Keywords