Optimization of contrast medium volume for abdominal CT in oncologic patients: prospective comparison between fixed and lean body weight-adapted dosing protocols

Abstract

Background

Patient body size represents the main determinant of parenchymal enhancement and by adjusting the contrast media (CM) dose to patient weight may be a more appropriate approach to avoid a patient over dosage of CM. To compare the performance of fixed-dose and lean body weight (LBW)-adapted contrast media dosing protocols, in terms of image quality and parenchymal enhancement.

Results

One-hundred cancer patients undergoing multiphasic abdominal CT were prospectively enrolled in this multicentric study and randomly divided in two groups: patients in fixed-dose group (n = 50) received 120 mL of CM while in LBW group (n = 50) the amount of CM was computed according to the patient’s LBW. LBW protocol group received a significantly lower amount of CM (103.47 ± 17.65 mL vs. 120.00 ± 0.00 mL, p < 0.001). Arterial kidney signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) and pancreatic CNR were significantly higher in LBW group (all p ≤ 0.004). LBW group provided significantly higher arterial liver, kidney, and pancreatic contrast enhancement index (CEI) and portal venous phase kidney CEI (all p ≤ 0.002). Significantly lower portal vein SNR and CNR were observed in LBW-Group (all p ≤ 0.020).

Conclusions

LBW-adapted CM administration for abdominal CT reduces the volume of injected CM and improves both image quality and parenchymal enhancement.

Key points

  • Lean body weight (LBW) group received significantly lower amount of contrast media.

  • Higher arterial contrast enhancement index (CEI) and kidney portal-venous CEI in LBW-group.

  • LBW protocol may be considered for routine to improve CT performance.

Background

Contrast-enhanced CT is the imaging modality of choice for the study of the abdomen in a large spectrum of diseases. Patient body size is known to be a major determinant of parenchymal enhancement, and it has been shown that adjusting the contrast media (CM) dose to patient weight may be a more appropriate approach than administering a fixed dose of CM, allowing to reduce inter-patient variability and unnecessary healthcare costs related to CM over dosage [1,2,3,4,5,6]. However, there are no general recommendations or established guidelines regarding the CM dosing method necessary to optimize parenchymal enhancement for abdominopelvic CT studies, and fixed-dose CM injection protocols are still commonly used for clinical practice in many institutions [7, 8].

Due this lack of standardization, multiple weight-adapted CM dosing regimens have been proposed, based on a large variety of body size indexes, including total body weight (TBW), body mass index (BMI), lean body weight (LBW), and body surface area (BSA), but no consensus on has been reached so far [9,10,11,12,13].

Dosing CM according to patient TBW does not require complicated calculations and, given its quickness and ease of use, can be readily implemented in the daily routine; nonetheless, it fails to take into account differences in body composition. In particular, TBW-adapted regimens may lead to an overestimation of CM volume in overweight and obese patients, in which scarcely perfused adipose tissue contributes for a considerable proportion of body weight [14, 15]. Differences in body fat percentage between men and women may also result in excessive amount of CM being administered to women, when dosing CM according to TBW [16, 17].

Promising results have been reported by previous studies investigating CM dosing protocols based on LBW rather than TBW, which resulted in a reduced inter-patient variability of parenchymal and vascular enhancement [16, 18]. Tailoring CM volume to patient LBW also demonstrated to better correlate with aortic and liver enhancement compared to TBW, BSA, or blood volume [19,20,21,22]. However, the effectiveness of LBW-adapted dosing protocols has been mostly investigated exclusively on Asian population, with many studies not assessing for differences in subjective image quality.

Therefore, our aim was to perform a prospective multicenter study to compare the performance of fixed-dose and LBW-adapted CM dosing protocols for multiphasic abdominal CT, in terms of image quality and parenchymal enhancement.

Methods

Study population

This prospective randomized multicenter study was conducted at two centers (Sant'Andrea University Hospital, Rome, Italy; and ICOT Hospital, Latina, Italy) and was approved by the Institutional Review Board of both participating institutions. Written informed consent was obtained from all patients and the study was compliant with the Health Insurance Portability and Accountability Act. From November 2018 to March 2019 oncologic patients clinically referred for multiphasic contrast-enhanced abdominal CT were prospectively enrolled. Exclusion criteria were as follows: (a) age < 18 years; (b) history of allergic reactions to iodinated CM; (c) kidney failure (estimated glomerular filtration rate < 30 mL/min/1.73 m2); (d) pregnancy; (e) deviation from injection and acquisition protocol and (f) presence of image artifacts deemed to impair quantitative measurement. Data on patient age, gender, height, primary cancer, TBW, LBW and BMI were recorded for all participants.

Contrast media injection protocol

Iomeprol with an iodine concentration of 350 mgI/mL (Iomeron 350; Bracco Imaging, Italy), was intravenously injected at a fixed flow rate of 3.0 mL/s (Iodine delivery rate of 1.05 gI/s) through an 18-gauge antecubital access by using an automated dual-syringe power injector (Stellant D; Medrad Inc, Warrendale, PA) and followed by a 50 mL saline chaser administered at the same flow rate.

Patients were randomly assigned into one of two CM dosing protocols using a randomization list on a 1:1 basis. Patients in the fixed dose protocol group received a fixed CM dose of 120 mL [42 g of iodine (gI) ]. Patients in the LBW protocol group received 0.7 gI per kg of LBW [10]. The resulting value was then divided by CM concentration (mgI/mL) to obtain the correct CM volume to be administered, as follows:

$$\mathrm{CM}\, \mathrm{volume }\left(\mathrm{mL}\right)=\frac{0.7 \cdot \mathrm{LBW}}{350} \cdot 1000$$

LBW was calculated by using James formula as follows [23]:

$$\begin{aligned}\mathrm{LBW}\,\mathrm{men}&=\left(1.10 \cdot {W}\right)-128 \cdot \left[\frac{{W}^{2}}{{(100 \cdot H)}^{2}}\right]\\ \mathrm{LBW}\,\mathrm{women}&=\left(1.07 \cdot {W}\right)-148 \cdot \left[\frac{{W}^{2}}{{(100 \cdot H)}^{2}}\right]\end{aligned}$$

where W is the patient weight in kilograms and H is the patient height in meters. For each CT examination, the volume of administered CM was recorded.

Scanning protocol and image reconstruction

Imaging was performed by using a 64-row multidetector CT scanner (Lightspeed VCT, GE Medical Systems, Waukesha, WI, USA) or with a 256-slice CT (Brilliance iCT 256, Philips Healthcare, The Netherlands), based on the reference centre. However, both in fixed dose protocol group and in the LBW protocol group the number of CT scans performed with two different CT scanners were equivalent, as reported in Table 1.

Table 1 Contrast media protocol and CT technical specifications

Scanning parameters for 64-row CT have been set as follows: tube voltage, 120 kVp; beam pitch, 1.375:1; detector configuration, 64 × 0.625 mm. A z-axis tube current modulation (Smart mA, GE Healthcare) was applied with a noise index of 28 (min/max tube current: 200/600 mAs).

For 256-slice CT, scanning parameters were: tube voltage, 120 kVp; beam pitch, 1; detector configuration, 2 × 128 rows × 0.625 mm; tube load from 200 to 600 mAs depending on automatic mA current modulation (Dose Right ACS, Philips Healthcare), with a Dose Right Index (DRI) of 22 and a Liver DRI of + 2.

All examinations were performed with patient in supine position and in a cranio-caudal direction, from the diaphragmatic dome to the pubic symphysis. The scan delay was determined using a bolus-tracking software program (SmartPrep for GE Healthcare, BolusPro for Philips Healthcare), with the placement of a 100 HU-threshold region-of-interest (ROI) within the abdominal aorta at the level of the celiac tripod. All patients in the two centres underwent a hepatic dynamic scan, including a pre-enhanced phase, a late arterial phase acquired 16 s after reaching the threshold, and a portal venous phase acquired 70 s after reaching the threshold.

Image datasets were reconstructed at the CT scanner console with the following parameters: slice thickness of 1.25 mm and spacing of 1.25 mm. Iterative reconstruction programs, “ASiR 40%; GE Healthcare” and “iDose4—level 2; Philips Healthcare”, were applied, as vendors suggested.

Objective image quality analysis

In both centres, a reader with at least 3 years of experience (DDS) in abdominal radiology analysed all images. Attenuation measurements in HU and standard deviation (SD) values, were obtained in axial scans, by positioning a circular ROI of approximately 1 cm2 in the liver (segment II, IVa and VII), renal cortex of both kidneys, pancreas, and left psoas muscle in both arterial and portal venous phase. Supplementary ROIs were placed in the suprarenal abdominal aorta on arterial phase images and in the portal vein on portal venous phase images. All measurements were performed three times at the same levels and then averaged to ensure consistency. The mean of the 3 averaged liver ROIs was obtained to define the liver HU; kidney HU was defined as the mean of right and left kidney averaged ROIs. Image noise was defined as the SD measured in a circular ROI placed in subcutaneous fat tissue.

Arterial liver, pancreatic, kidney and aortic signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) as well as portal venous phase liver, pancreatic, kidney, and portal vein SNR and CNR were calculated as follows [12]:

$$\begin{aligned}\mathrm{SNR}&= \frac{\mathrm{HU}}{\mathrm{Noise}}\\ \mathrm{CNR }&= \frac{\mathrm{HU}-\mathrm{HU}\,\mathrm{muscle}}{\mathrm{Noise}}.\end{aligned}$$

Liver, kidney and pancreas contrast enhancement on both arterial and portal venous phase was quantified by calculating the contrast enhancement index (CEI) as follows [19]:

$${\text{CEI}} = {\text{HU}}\;{\text{enhanced}} - {\text{HU}}\,{\text{unenhanced}}{.}$$

Subjective image quality analysis

Subjective image quality was assessed independently by two experienced abdominal radiologists (M.Z. and N.P.) at each centre. All readers were unaware of which CM dosing protocol had been used. Image datasets were primarily displayed with standard window settings for evaluation of soft tissue (width: 400 HU; level: 40 HU). However, readers were allowed to freely adjust window width and level values according to their preferences. The enhancement of the liver, kidneys, and pancreas was rated by using a 5-point Likert scale as follows: 1 = very poor; 2 = poor; 3 = fair; 4 = good; 5 = excellent [20].

Statistical analysis

Statistical analysis was performed by using the MedCalc5 Statistical Software version 17.9.7 (MedCalc Software bvba, Ostend, Belgium; http://www.medcalc.org; 2017). Continuous variables were expressed as mean ± SD or mean with ranges. Ordinal variables were expressed as median with ranges. Patient characteristics (BMI, age, and CM dose expressed in gI and mL), SNR, CNR, and CEI, were compared between the two groups. The Kolmogorov–Smirnov test was used to assess data distribution. For normally distributed data, Student’s t test was applied. In case of non-normally distributed data, Mann–Whitney U test was performed. The x2 test was used to calculate differences in gender and type of primary cancer between the two study groups.

Cohen’s Kappa test was used to evaluate inter-reader agreement for subjective image quality assessment using the following coefficients: κ ≤ 0.20, poor agreement; κ = 0.21–0.40, fair agreement; κ = 0.41–0.60, moderate agreement; κ = 0.61–0.80, good agreement; and κ = 0.81–1.0, excellent agreement [24]. A p value ≤ 0.05 was considered statistically significant.

Results

Study population

One-hundred-ten consecutive patients were prospectively included in the study, 10 individuals were excluded due to previous allergic reactions to iodinated CM (n = 2), kidney failure (n = 3), deviation from injection protocol (n = 2) and the presence of artifacts (n = 3). Therefore, 100 patients were finally enrolled in the study population: 50 patients in the fixed dose protocol group and 50 patients in the LBW protocol group. No significant differences were observed between the two groups in terms of patient age, gender, BMI, and LBW (all p ≥ 0.111). Detailed patient characteristics are reported in Table 2.

Table 2 Patients characteristics

Contrast media dose

The administered CM dose was significantly lower in LBW protocol group compared to fixed dose protocol group, both in terms of gI (36.2 ± 6.2 gI vs. 42.0 ± 0 gI, p < 0.001) and corresponding volume (103.4 ± 17.6 mL vs. 120.0 ± 0 mL, p < 0.001), as shown in Table 2. Within LBW protocol group, 7 patients (13.7%) received a CM dose greater than 120 mL (mean CM dose: 132.29 ± 8.34 mL, CM dose range: 124–144 mL).

Objective image quality analysis

LBW protocol group accounted for significantly higher liver arterial CEI (23.02 ± 10.84 vs. 16.10 ± 9.65; p < 0.001), pancreatic arterial CNR (3.85 ± 2.04 vs. 2.92 ± 1.76; p = 0.004), pancreatic arterial CEI (66.45 ± 19.78 vs. 50.66 ± 17.29; p < 0.001), kidney arterial SNR (8.19 ± 2.61 vs. 7.38 ± 4.17; p = 0.016), and kidney arterial CNR (9.60 ± 3.55 vs. 7.90 ± 4.10; p = 0.003). Kidney CEI was superior in LBW protocol group in comparison with fixed dose group for both arterial (145.00 ± 42.53 vs. 122.17 ± 35.68; p = 0.002) and portal venous phase (103.56 ± 30.35 vs. 78.97 ± 58.75; p < 0.001).

Significantly lower portal vein SNR (9.33 ± 3.37 vs. 10.42 ± 3.01; p = 0.015) and CNR (7.57 ± 2.31 vs. 8.92 ± 3.31; p = 0.020) were observed in LBW protocol group compared to fixed dose group.

No significant differences were observed in any of the remaining image quality parameters evaluated on both arterial and portal venous phases (all p ≥ 0.079). Results of objective image quality analysis are summarized in Table 3 and Figs. 1, 2, and 3.

Table 3 Results of objective image quality analysis
Fig. 1
figure 1

Box-and-whisker plots show average signal-to-noise ratios (SNR) of liver, pancreas, kidney, aorta, and portal vein on both arterial phase (AP) and portal venous phase (PVP), compared between fixed dose group (fixed) and lean body weight group (LBW). Boxes represent 25th and 75th percentile, horizontal lines 50th percentile (median), and whiskers minimum and maximum values

Fig. 2
figure 2

Box-and-whisker plots show average contrast-to-noise ratios (CNR) of liver, pancreas, kidney, aorta, and portal vein on both arterial phase (AP) and portal venous phase (PVP), compared between fixed dose group (fixed) and lean body weight group (LBW). Boxes represent 25th and 75th percentile, horizontal lines 50th percentile (median), and whiskers minimum and maximum values

Fig. 3
figure 3

Box-and-whisker plots show average contrast enhancement index (CEI) of liver, pancreas, and kidney on both arterial phase (AP) and portal venous phase (PVP), compared between fixed dose group (fixed) and lean body weight group (LBW). Boxes represent 25th and 75th percentile, horizontal lines 50th percentile (median), and whiskers minimum and maximum values

Subjective image quality analysis

Detailed results are reported in Table 4. Subjective image quality assessment returned overall comparable results between the two group. LBW protocol group provided slightly higher contrast enhancement ratings compared to fixed dose protocol group for portal venous phase liver enhancement, arterial pancreatic enhancement, and both arterial and portal venous phase kidney enhancement, although all differences were not statistically significant (all p ≥ 0.213). No patients scored very poor or poor enhancement in both groups.

Table 4 Results of subjective image quality analysis

Overall inter-rater agreement was excellent for both fixed dose protocol (κ = 0.81 (95% CI 0.76–0.87)) and LBW protocol (κ = 0.87 (95% CI 0.81–0.93)) groups.

Discussion

The aim of our study was to compare the performance of fixed-dose and LBW-adapted CM dosing protocols for multiphasic abdominal CT, in terms of image quality and parenchymal enhancement in oncologic patients. Our results showed that tailoring the CM volume according to LBW provides higher parenchymal enhancement of abdominal organs than using a fixed CM volume. Moreover, comparable or higher SNR and CNR were obtained in both arterial and portal venous phases despite the lower CM volume injected in the LBW tailored group.

Weight-adapted strategies tailored on patient’s TBW have been advocated for CM administration in multiphasic abdominopelvic CT, in order to overcome the drawbacks reported with fixed-dose protocols, namely the tendency to overestimation of CM amount in smaller patients or underdosage in heavier patients [5]. However, this technique may result in overeating of CM amount when applied to underweight individuals [25]. On the other end, adjusting the CM dose based on TBW may results in unnecessary elevated volumes of CM to be injected in overweight individuals, as a considerable fraction of their TBW is composed of adipose tissue, which has a negligible impact on solid organs’ enhancement [18, 26]. Parenchymal CM concentrations strongly depend on extracellular compartment volume since, after intravenous injection, CM promptly distributes from intravascular space to parenchymal extracellular space, without permeating into the intracellular space [2].

Among different body indexes, LBW has been demonstrated to better correlate with plasma and extracellular space volume [27, 28]. Therefore, dosing CM according to patients’ LBW accounted for differences in body composition and allowed to achieve consistent parenchymal enhancement even at reduced iodine load, by excluding from calculation of CM dose the irrelevant contribution of poorly perfused adipose tissue. Our results are in agreement with and corroborated those from previous studies that found optimal correlation between LBW and parenchymal enhancement, with improved per-patient uniformity [16, 19, 20].

Moreover, results obtained showed how CEI was higher in LBW protocol for liver and pancreas in arterial phase and for kidney in both arterial and portal venous phases. Despite these results might seem counterintuitive due to parenchymal enhancement dependency to iodine load, both CNR and SNR have no significant differences in liver, a solid organ usually considered for quantitative analysis [3, 5, 6, 9]. Regarding kidney and pancreas discrepancies, we hypothesized that measurements might be affected by the possible reduced dimension of kidney cortex while pancreas has not been studied in the specific late arterial pancreatic phase but in a late arterial phase tailored for multiphasic CT protocol for oncologic follow-up purpose.

Although it is beyond the scope of the study, we also observed that LBW-adapted protocol performed better in terms of parenchymal rather than vascular enhancement with reduced portal vein CNR and SNR in LBW-adapted protocol compared to fixed dose protocol. This result may be explained by the larger average CM volume administered to patients in the fixed dose protocol group. Indeed, vascular enhancement has been proven to be directly proportional to the injected CM volume, when injection rate and duration are maintained stable [2, 29]. Nonetheless, our LBW-adapted protocol is intended to be applied to multiphasic CT examinations in oncological setting, in which the main goal is tumor detection and response assessment rather than obtaining a robust angiographic study. Anyhow, mean aortic artery and portal vein attenuations in the LBW protocol group exceeded by far the minimal enhancement deemed diagnostic in discriminating between vessels and lymph nodes [30].

In our study, patients in the LBW-protocol group were given a significantly lower average amount of CM compared to those assigned to the fixed dose protocol. A reduced CM volume is especially beneficial for patients with cancer, who require multiple contrast-enhanced CT examinations to assess for disease progression and monitor the response to therapy. Moreover, it has been shown that oncologic patients are at high risk for developing acute kidney adverse events following iodinated CM administration, given the increased prevalence of associated risk factors, such as pre-existing kidney insufficiency, advanced age, dehydration, and concurrent nephrotoxic chemotherapeutic regimens [31, 32].

Lowering the average dose of administered CM should also be advisable in order to avoid unnecessary healthcare costs. Although a cost-effectiveness analysis was beyond the aim of our investigation, the potential cost savings achievable by the implementation of the present LBW-adapted CM dosing protocol are suggested by the results from previous studies that reported a remarkable cost reduction when a weight-adapted rather than a fixed-dose protocol is used [1, 7, 25].

There are several limitations of the present study that should be mentioned. First, as we assessed only a single type of CM (i.e. iodine concentration of 350 mgI/mL), additional studies would be required to demonstrate the reproducibility of our result when using CM with different iodine concentrations. Second, we solely evaluated the performance of LBW-adapted protocol in terms of image quality, further analysis to determine also the effects on lesions’ conspicuity and diagnostic accuracy was not investigated. Third, we calculated the LBW of each patient by using James formula; however, several prediction formulas for LBW have been previously reported and their application may yield results different from those observed in our study [27, 28]. However, the James formula is frequently used for estimating the CM dose and, although the use of Boer formula has been recommended in patients with a high BMI [33], no significant differences in objective image quality have been reported between these two formulas, when applied in a range of BMI as that observed in our population [14]. Furthermore, we determined the CM volume to be administered for LBW protocol group according to the calculated LBW rather than measured LBW, estimated by measuring patient fat body percentage with the aid of an analyzer scale [34]. Although the latter technique may have yielded a more accurate estimation, this approach would be technically impractical, while the use of calculated LBW supports the broad applicability of the present LBW-adapted protocol in the routine clinical practice. Lastly, although a lower iodine load achievable with LBW-adapted protocol could potentially reduce the risk of acute kidney failure in patients affected by renal chronic disease compared with a regular CM protocol, this aspect was not assessed in the study.

Conclusions

In conclusion, the results of this prospective randomized multicenter study demonstrate that dosing CM according to patient’s LBW rather than administering a fixed dose of CM allows for a significant reduction of the injected CM volume with no detrimental effects on image quality and parenchymal enhancement. Implementation of an LBW-adapted protocol should be considered in oncologic patients in order to reduce costs and minimize the risks of contrast-induced acute kidney adverse events.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

BMI:

Body mass index

BSA:

Body surface area

CEI:

Contrast enhancement index

CM:

Contrast media

CNR:

Contrast-to-noise ratio

CT:

Computed tomography

LBW:

Lean body weight

ROI:

Region-of-interest

SNR:

Signal-to-noise ratio

References

  1. Arana E, Marti-Bonmati L, Tobarra E, Sierra C (2009) Cost reduction in abdominal CT by weight-adjusted dose. Eur J Radiol 70:507–511

    Article  Google Scholar 

  2. Bae KT (2010) Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology 256:32–61

    Article  Google Scholar 

  3. Benbow M, Bull RK (2011) Simple weight-based contrast dosing for standardization of portal phase CT liver enhancement. Clin Radiol 66:940–944

    Article  CAS  Google Scholar 

  4. George AJ, Manghat NE, Hamilton MC (2016) Comparison between a fixed-dose contrast protocol and a weight-based contrast dosing protocol in abdominal CT. Clin Radiol 71:1314.e1311–1314.e1319

    Article  Google Scholar 

  5. Yamashita Y, Komohara Y, Takahashi M et al (2000) Abdominal helical CT: evaluation of optimal doses of intravenous contrast material—a prospective randomized study. Radiology 216:718–723

    Article  CAS  Google Scholar 

  6. Rengo M, Caruso D, De Cecco CN et al (2012) High concentration (400 mgI/mL) versus low concentration (320 mgI/mL) iodinated contrast media in multi detector computed tomography of the liver: a randomized, single centre, non-inferiority study. Eur J Radiol 81:3096–3101

    Article  Google Scholar 

  7. Davenport MS, Parikh KR, Mayo-Smith WW, Israel GM, Brown RK, Ellis JH (2017) Effect of Fixed-volume and weight-based dosing regimens on the cost and volume of administered iodinated contrast material at abdominal CT. J Am Coll Radiol 14:359–370

    Article  Google Scholar 

  8. Zanardo M, Doniselli FM, Esseridou A et al (2018) Abdominal CT: a radiologist-driven adjustment of the dose of iodinated contrast agent approaches a calculation per lean body weight. Eur Radiol Exp 2:41

    Article  Google Scholar 

  9. Kidoh M, Nakaura T, Oda S et al (2013) Contrast enhancement during hepatic computed tomography: effect of total body weight, height, body mass index, blood volume, lean body weight, and body surface area. J Comput Assist Tomogr 37:159–164

    Article  Google Scholar 

  10. Kondo H, Kanematsu M, Goshima S et al (2013) Body size indices to determine iodine mass with contrast-enhanced multi-detector computed tomography of the upper abdomen: does body surface area outperform total body weight or lean body weight? Eur Radiol 23:1855–1861

    Article  Google Scholar 

  11. Onishi H, Murakami T, Kim T et al (2011) Abdominal multi-detector row CT: effectiveness of determining contrast medium dose on basis of body surface area. Eur J Radiol 80:643–647

    Article  Google Scholar 

  12. Rengo M, Bellini D, De Cecco CN et al (2011) The optimal contrast media policy in CT of the liver. Part I: technical notes. Acta Radiol 52:467–472

    Article  Google Scholar 

  13. Rengo M, Bellini D, De Cecco CN et al (2011) The optimal contrast media policy in CT of the liver. Part II: clinical protocols. Acta Radiol 52:473–480

    Article  Google Scholar 

  14. Caruso D, De Santis D, Rivosecchi F et al (2018) Lean body weight-tailored iodinated contrast injection in obese patient: Boer versus James formula. Biomed Res Int 2018:8521893

    PubMed  PubMed Central  Google Scholar 

  15. Kondo H, Kanematsu M, Goshima S et al (2008) Abdominal multdetector CT in patients with varying body fat percentages: estimation of optimal contrast material dose. Radiology 249:872–877

    Article  Google Scholar 

  16. Peet K, Clarke SE, Costa AF (2018) Hepatic enhancement differences when dosing iodinated contrast media according to total versus lean body weight. Acta Radiol 60:807–814

    Article  Google Scholar 

  17. Tanaka J, Kozawa E, Inoue K, Okamoto Y, Toya M, Sato Y (2011) Should the dose of contrast medium be determined solely on the basis of body weight regardless of the patient’s sex? Jpn J Radiol 29:330–334

    Article  Google Scholar 

  18. Ho LM, Nelson RC, Delong DM (2007) Determining contrast medium dose and rate on basis of lean body weight: does this strategy improve patient-to-patient uniformity of hepatic enhancement during multi-detector row CT? Radiology 243:431–437

    Article  Google Scholar 

  19. Awai K, Kanematsu M, Kim T et al (2016) The optimal body size index with which to determine iodine dose for hepatic dynamic CT: a prospective multicenter study. Radiology 278:773–781

    Article  Google Scholar 

  20. Kondo H, Kanematsu M, Goshima S et al (2010) Body size indexes for optimizing iodine dose for aortic and hepatic enhancement at multidetector CT: comparison of total body weight, lean body weight, and blood volume. Radiology 254:163–169

    Article  Google Scholar 

  21. Caruso D, Eid M, Schoepf UJ et al (2017) Optimizing contrast media injection protocols in computed tomography angiography at different tube voltages: evaluation in a circulation phantom. J Comput Assist Tomogr 41:804–810

    Article  Google Scholar 

  22. De Santis D, Caruso D, Schoepf UJ et al (2018) Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom. Eur Radiol 28:3473–3481

    Article  Google Scholar 

  23. James WPT (1977) Research on obesity*. Nutr Bull 4:187–190

    Article  Google Scholar 

  24. Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174

    Article  CAS  Google Scholar 

  25. Megibow AJ, Jacob G, Heiken JP et al (2001) Quantitative and qualitative evaluation of volume of low osmolality contrast medium needed for routine helical abdominal CT. AJR Am J Roentgenol 176:583–589

    Article  CAS  Google Scholar 

  26. Kondo H, Kanematsu M, Goshima S et al (2011) Aortic and hepatic enhancement at multidetector CT: evaluation of optimal iodine dose determined by lean body weight. Eur J Radiol 80:e273–e277

    Article  Google Scholar 

  27. Boer P (1984) Estimated lean body mass as an index for normalization of body fluid volumes in humans. Am J Physiol 247:F632–F636

    CAS  PubMed  Google Scholar 

  28. Peters AM, Snelling HL, Glass DM, Love S, Bird NJ (2010) Estimated lean body mass is more appropriate than body surface area for scaling glomerular filtration rate and extracellular fluid volume. Nephron Clin Pract 116:c75-80

    Article  Google Scholar 

  29. Fleischmann D, Kamaya A (2009) Optimal vascular and parenchymal contrast enhancement: the current state of the art. Radiol Clin N Am 47:13–26

    Article  Google Scholar 

  30. Awai K, Imuta M, Utsunomiya D et al (2004) Contrast enhancement for whole-body screening using multidetector row helical CT: comparison between uniphasic and biphasic injection protocols. Radiat Med 22:303–309

    PubMed  Google Scholar 

  31. Ng CS, Kalva SP, Gunnarsson C, Ryan MP, Baker ER, Mehta RL (2018) Risk of renal events following intravenous iodinated contrast material administration among inpatients admitted with cancer a retrospective hospital claims analysis. Cancer Imaging 18:30

    Article  Google Scholar 

  32. Cicin I, Erdogan B, Gulsen E et al (2014) Incidence of contrast-induced nephropathy in hospitalised patients with cancer. Eur Radiol 24:184–190

    Article  Google Scholar 

  33. Nyman U (2016) James lean body weight formula is not appropriate for determining CT Contrast media dose in patients with high body mass index. Radiology 278:956–957

    Article  Google Scholar 

  34. Rengo M, Bellini D, Businaro R et al (2017) MDCT of the liver in obese patients: evaluation of a different method to optimize iodine dose. Abdom Radiol (NY) 42:2420–2427

    Article  Google Scholar 

Download references

Funding

The authors received no specific funding for this work.

Author information

Authors and Affiliations

Authors

Contributions

All authors read and approved the final manuscript. DC design of the work, analysis, interpretation of data, revised the work; ER acquisition, interpretation of data, draft the work; NP conception of the work, acquisition, draft the work; MR interpretation of data, revised the work; DB conception of the work, revised the work; GM draft the work; BB acquisition; EL interpretation of data; MZ draft the work; MP interpretation of data; DDS draft the work; EI conception of the work; PA design of the work, interpretation of data; IC design of the work, interpretation of data, revised the work; AL conception of the work, analysis, interpretation of data, revised the work.

Corresponding author

Correspondence to Andrea Laghi.

Ethics declarations

Ethics approval and consent to participate

This prospective randomized multicenter study was conducted at two centers (Sant'Andrea University Hospital; and ICOT Hospital) and was approved by the Institutional Review Board of both participating institutions.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Caruso, D., Rosati, E., Panvini, N. et al. Optimization of contrast medium volume for abdominal CT in oncologic patients: prospective comparison between fixed and lean body weight-adapted dosing protocols. Insights Imaging 12, 40 (2021). https://doi.org/10.1186/s13244-021-00980-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13244-021-00980-0

Keywords