Skip to main content

Table 4 Intra- and inter-observer variability and the comparison between 2D and 3D k-t BLAST in terms of global LV volume parameters and reporting time. Parameters are expressed as mean ± SD; p value Student’s paired test; r value Pearson’s correlation; 95% LA limits of agreement with Bland-Altman analysis

From: Magnetic resonance assessment of left ventricular volumes and mass using a single-breath-hold 3D k-t BLAST cine b-SSFP in comparison with multiple-breath-hold 2D cine b-SSFP

 

Intra-observer

Inter-observer

3D k-t BLAST

1st analysis

2nd analysis

p value

r value

95% LA

1st observer

2nd observer

p value

r value

95% LA

EF (%)

43 ± 15

43 ± 15

>0.05

0.99

1.75; -2.23

43 ± 15

41 ± 14

<0.05

0.96

10.2; -6.5

ED wall mass (g)

81 ± 24

82 ± 24

>0.05

0.99

3.8; -4.1

81 ± 24

88 ± 24

<0.05

0.93

10.6; -24.3

Report time (min)

29 ± 7

29 ± 7

>0.05

0.99

2.5; -2.1

29 ± 7

32 ± 7

<0.05

0.89

3.1; -10.2

2D

1st analysis

2nd analysis

   

1st observer

2nd observer

   

EF (%)

48 ± 15

48 ± 15

>0.05

0.99

2.3; -1.7

48 ± 15

46 ± 14

<0.05

0.96

9.8; -5.9

ED wall mass (g)

72 ± 23

72 ± 23

>0.05

0.99

6.8; -4.9

72 ± 23

76 ± 23

<0.05

0.92

13.9; -20.5

Report time (min)

8 ± 3

8 ± 2

>0.05

0.90

2.2; -2.7

8 ± 3

9 ± 3

<0.05

0.64

3.2; -6.4

2D vs 3D k-t BLAST

2D

3D k-t BLAST

   

2D

3D k-t BLAST

   
 

2nd analysis

2nd analysis

   

1st observer

2nd observer

   

EF (%)

48 ± 15

43 ± 15

<0.05

0.99

6.56; 2.75

46 ± 14

41 ± 14

<0.05

0.99

7.9; 2.2

ED wall mass (g)

72 ± 23

82 ± 24

<0.05

0.98

0.4; -20.4

76 ± 23

88 ± 24

<0.05

0.96

-0.5; -24.7

Report time (min)

8 ± 2

29 ± 7

<0.05

0.25

-6.9; -34.4

9 ± 3

29 ± 7

<0.05

0.16

-4.8; -34.1