Skip to main content

Table 3 Comparison of global LV parameters calculated with 2D and 3D k-t BLAST sequences. Parameters are expressed as mean ± SD; p value Student’s paired test; r value Pearson’s correlation; 95% LA limits of agreement with Bland-Altman analysis

From: Magnetic resonance assessment of left ventricular volumes and mass using a single-breath-hold 3D k-t BLAST cine b-SSFP in comparison with multiple-breath-hold 2D cine b-SSFP

LV Parameters

2D

3D k-t BLAST

p value

r value

95% LA

EDV (ml)

134 ± 49

129 ± 44

<0.05

0.98

-16.4; 26.6

ESV (ml)

73 ± 50

77 ± 47

<0.05

0.99

-16.7; 9.7

SV (ml)

60 ± 21

52 ± 19

<0.05

0.96

-2.7; 19.9

EF (%)

48 ± 15

43 ± 15

<0.01

0.99

2.7; 7.6

CO (l/min)

3.9 ± 1.3

3.3 ± 1.3

<0.05

0.95

-0.20; 1.34

ED wall mass (g)

72 ± 23

81 ± 24

<0.05

0.97

-19.7; 1.6

  1. LV left ventricle, EDV end diastolic volume, ESV end systolic volume, SV stroke volume, EF ejection fraction, CO cardiac output, ED wall mass end diastolic wall mass