Skip to main content

Table 5 Summary of the intra- and interrater correlation of quality scores for each quantitative metric. Metrics are identified according to the letters assigned in Table 2

From: Image quality in whole-body MRI using the MY-RADS protocol in a prospective multi-centre multiple myeloma study

Intra-rater

Metric

A

B

C1

C2

D

E1

E2

F1

F2

G

 ICC

0.91

0.17

0.95

0.88

0.83

0.74

0.32

0.59

0.66

0.07

 p

 < .001

.282

 < .001

.006

 < .001

.004

.138

.019

.016

.372

 Mean bias

 − 0.06

0.16

0.83

 − 1.23

 − 0.05

0.10

3.29

0.38

0.27

 − 0.80

Inter-rater

Metric

A

B

C1

C2

D

E1

E2

F1

F2

G

 ICC

0.51

0.01

0.85

0.40

0.36

0.92

0.90

0.65

0.08

0.08

 p

.032

.489

.005

.211

.131

 < .001

 < .001

.007

.413

.413

 Mean bias

 − 0.77

 − 0.03

 − 1.83

1.98

 − 0.06

0.00

 − 0.52

0.37

 − 0.65

 − 0.65

  1. A SNR B anterior thoracic signal loss, C1 susceptibility: number of slices, C2 susceptibility: length, D fat suppression, E1 ghosting (brain), E2 ghosting (pelvis), F1 geometric distortion (brain), F2 geometric distortion (pelvis), G eddy current distortion