Skip to main content

Table 4 The model was used to predict the radiological DWI scan quality using all ten quantitative metrics. Metrics that were found to be statistically significant predictors are indicated with an asterisk

From: Image quality in whole-body MRI using the MY-RADS protocol in a prospective multi-centre multiple myeloma study

Metric

Coefficient (β)

p-value

Odds ratio

Odds ratio 95% CI

SNR

*

 − 0.483

.032

0.62

0.40–0.96

Anterior thoracic signal loss

*

 − 0.716

.002

0.49

0.31–0.76

Susceptibility artefact—total no. slices

 

0.311

.490

1.36

0.56–3.31

Susceptibility artefact—total length

 

 − 0.616

.176

0.54

0.22–1.32

Eddy current distortion

 

0.357

.148

1.43

0.88–2.32

Fat suppression

 

 − 0.329

.132

0.72

0.47–1.10

Ghosting—brain

 

 − 0.469

.060

0.63

0.38–1.02

Ghosting—pelvis

 

 − 0.457

.087

0.63

0.38–1.07

Distortion—brain

*

 − 0.536

.019

0.59

0.37–0.92

Distortion—pelvis

 

 − 0.222

.341

0.80

0.51–1.26

  1. *Metrics that were found to be statistically significant predictors are indicated with an asterisk