Skip to main content

Table 2 Imaging findings of mechanical circulatory support device complications

From: CT and chest radiography in evaluation of mechanical circulatory support devices for acute heart failure

Complication

CXR

Echo/US

CT

Malposition

Markers not at expected landmarks

Change in positionĀ compared to prior imaging

Markers not at expected landmarks

Change in positionĀ compared to prior imaging

Markers not at expected landmarks

Change in positionĀ compared to prior imaging

Hematoma

New mediastinal widening or pleural parenchymal opacities

Hemopericardium with/without tamponade

High attenuation fluid collection. Active contrast extravasation on contrast enhanced CT

Cardiovascular perforation

Tip beyond the cardio mediastinal contour

Inability to dislodge the device

Tip in the mediastinum with associated hemopericardium or hemothorax

Thrombus

ā€“

Echogenic lesion abutting the device

Acute thrombus can be hyperdense to blood pool on non-contrast CT

Chronic Thrombus can have calcifications

Thrombus is hypodense to myocardium on contrast enhanced CT

Embolus

ā€“

Decreased/absent peripheral arterial flow

Signs of infarcts/ischemia due to distal emboli