Skip to main content
Fig. 3 | Insights into Imaging

Fig. 3

From: CT and chest radiography in evaluation of mechanical circulatory support devices for acute heart failure

Fig. 3

Complication from low IABP. AP chest radiograph (a) of a patient with an intra-aortic balloon pump. The upperĀ radiopaque marker is below the carina (white arrow) which suggests a low position. Sagittal image (b) of a contrast enhanced chest and abdomen CT at the level of the celiac trunk and superior mesenteric artery (SMA) origin demonstrates high grade obstruction of the SMA and celiac trunkĀ ostia by the distended IABP. Axial CT image (c) identifies complication from hypoperfusion: pneumatosis and contained perforation of the cecum (white arrow), due to the hypoperfusion from the high grade celiac trunk andĀ SMA ostialĀ occlusion. (IABPā€”Intra-aortic balloon pump)

Back to article page