Skip to main content

Table 1 DTI acquisition parameters

From: Inter-vendor and inter-observer reliability of diffusion tensor imaging in the musculoskeletal system: a multiscanner MR study

 

GE

Philips

Siemens

DTI parameters (Thigh)

 Coil

16 channels

32 channels

18 channels

 Directions

12

15

12

 b-value (s/mm2)

400

400

400

 TR (ms)

7350

7500

7500

 TE (ms)

Minimum

72

83.0

 Matrix

128 × 128

128 × 128

128 × 128

 Parallel factor

2 (asset)

2 (sense)

2 (grappa)

 Voxel size (mm)

3.1 × 3.1 × 4.0

3.03 × 2.97 × 4.0

3.1 × 3.1 × 4.0

 Acquisition time (minutes:seconds)

3.18

4.07

3.32

DTI parameters (leg)

   

 Coil

16 channels

32 channels

18 channels

 Directions

12

12

12

 b-value (s/mm2)

400

400

400

 TR (ms)

7950

7500

7500

 TE (ms)

Minimum

66

83

 Matrix

128 × 128

128 × 128

128 × 128

 Parallel factor

2 (asset)

2 (sense)

2 (grappa)

 Voxel size (mm)

2.3 × 2.3 × 4.0

2.3 × 2.3 × 4.0

2.3 × 2.3 × 4.0

 Acquisition time (minutes:seconds)

3.35

4.07

3.32

DTI parameters (arm)

   

 Coil

16 channels

8 channels

18 channels

 Directions

12

12

12

 b-value (s/mm2)

400.0

400

400

 TR (ms)

8600

7837

7900

 TE (ms)

Minimum

79

83

 Matrix

122 × 122

122 × 122

122 × 122

 Parallel factor

2 (asset)

2 (sense)

2 (grappa)

 Voxel size (mm)

1.6 × 1.6 × 4.0

1.6 × 1.6 × 4.0

1.6 × 1.6 × 4.0

 Acquisition time (minutes: seconds)

3.52

4.18

3.43