Table 3 Factors affecting protocol delivery at each site

From: Implementation of Whole-Body MRI (MY-RADS) within the OPTIMUM/MUKnine multi-centre clinical trial for patients with myeloma

Site

Protocol

Hardware limitation

Software limitation

Local preference (protocol changes from Table 2)

1

A

Ok

Ok

–

2

A

Ok

Ok

DWI: fewer averages (2,2,4) to reduce scan time; larger FOV (450 mm) to better cover the arms

3

B

Ok

No station composing

–

4

A

Ok

No station composing

–

5

C

Ok

Ok

–

6

D

Ok

No station composing (DWI only)

–

7

A

Ok

Ok

–

8

C

Ok

Ok

–

9

A

Low gradients (33 mT/m at 125 T/m/s vs. 45 mT/m at 200 T/m/s)

Ok

DWI: number of averages for b = 900 mm2/s was increased to 7 (from 6) to compensate for lower SNR (longer TE; 81 ms) due to lower gradient strength

10

C

Ok

Ok

–

11

A

Ok

Ok

DWI: fewer averages (2,2,4) to reduce scan time; larger FOV (455 mm) to better cover the arms

12

A

Ok

Ok

DWI: only 2 b-values (50, 900; omit b = 600 mm2/s) and fewer averages (2,5) to reduce scan time

  1. FOV field of view; SNR signal-to-noise ratio