Table 1 A summary of previous findings on potential feature exploration based on 18F-FDG PET/CT image of lung diseases

From: A review on advances in 18F-FDG PET/CT radiomics standardisation and application in lung disease management

No.

Dose (MBq)

Scanner type

Scan time (min)

Respiratory gating

Reconstruction algorithm

Matrix size

PSF Modelling

Time of flight (TOF)

Image processing filter

Type of segmentation

Type of quantisation

Tumour size (cm3)

REF

1

3.70–4.81

Discovery ST4, GE Healthcare

16–12

No

OSEM

CT: 512 × 512

No

No

Exponential filter, square filter, square root filter, logarithm filter and wavelet decomposition

Manual

–

–

[42]

2

VD: 370 FD: 400

Validation DS: VCT-XT-Discovery, GE-Healthcare Feasibility DS: GE Discovery ST PETCT system (GE Healthcare, Waukesha, WI, USA)

–

No

OSEM

–

No

No

None

Automated

–

–

[50]

3

210–620

Discovery ST, GE Healthcare, Waukesha, WI)

–

No

OSEM

–

No

No

–

Manual

Equal-probability and Lloyd–Max quantisation

–

[55]

4

220 – 690

Philips GE PET/CT scanner (Philips Medical Systems, USA)

–

No

RAMLA

–

No

No

5-mm full-width at half-maximum Gaussian

Semi-automated

64 grey-level quantisation

–

[57]

5

350–550

Siemens Biograph 6 LSO (Siemens, Erlangen Germany) or a General Electric Discovery 690 (General Electric Healthcare, Waukesha, WI, USA)

–

No

Iterative, TOF, sharp IR

PET: 128 × 128, CT: 512 × 512

No

Yes

None

Semi-automated

PET: 64 bins from 0 to 25,CT: 400 bins from − 1000 to 3000

1.64 ± 0.78

[77]

6

340–450

 

24–28

No

OSEM

CT: 512 × 512 128 × 128

No

No

None

Manual

256 bins quantisation

 < 3

[80]

7

370

ECAT EXACT 47 scanner (CTI/ Siemens, Munich, Germany)

–

No

OSEM

–

No

No

–

–

–

–

[86]

8

350–550

Siemens Biograph 6 LSO (Siemens, Erlangen, Germany), General Electric Discovery 690 (General Electric Healthcare, Waukesha, WI, USA)

–

No

Iterative, TOF, sharp IR

PET: 128 × 128, 256 × 256 CT: 512 × 512

No

Yes

–

Semi-automatic

Tonal discretisation (64 bins)

–

[94]

9

–

GE I PET/CT scanner (Philips),

–

No

RAMLA

–

No

No

5-mm full-width-at-half-maximum Gaussian

Automatic

–

–

[117]

10

370–555

General Electric Medical Systems, Waukesha, WI

35

Yes

OSEM

128 × 128

No

No

–

Automatic

–

–

[119]

11

618–814

Siemens Biograph PET/CT scanner (Siemens AG, Erlangen, Germany)

Non-gated: 18–35 Gated: 20–30

Yes

3D PET: OSEM, 4D PET: OSEM

No-gated: 168 × 168 Gated: 256 × 256

No

No

Non-gated: 7 mm full-width half- maximum Gaussian Gated: 5 mm full-width half-maximum Gaussian

–

32 discrete values quantisation

–

[122]

12

223–690

GE I PET/CT scanner (Philips)

–

No

RAMLA

–

No

No

5 mm in full width at half maximum Gaussian

Automatic

–

–

[138]

13

550

Philips GE I PET/CT, Philips Health Care, Cleveland, Ohio, USA

–

No

RAMLA

PET: 144 × 144, CT: 512 × 512

No

No

–

Manual

–

0.7–5.8

[6]

14

150–310

Discovery MI, GE Healthcare

15–20

No

OSEM BSREM

256 × 256

Yes

Yes

6.4-mm Gaussian filter with time-of-flight and PSF, and BSREM with a beta-value of 450

Manual

–

 ≤ 2

[71]

15

555

GE discovery LS 4 PET/CT scanner

–

No

OSEM

128 × 128

No

No

8 mm full-width at half-maximum Gaussian

Manual

256-bin discretisation

–

[76]

16

350 – 450

Biograph 16 Siemens Medical Solutions

20–25

No

OSEM

PET: 128 × 128, CT: 512 × 512

No

No

5 mm full-width at half-maximum Gaussian

Manual

–

1.7–6.8

[81]

17

–

Biograph 16 PET/CT scanner

18–21

No

OSEM

164 × 164

No

No

–

Manual

64 bins quantisation

1.7 – 6.8

[93]

18

270–410

Biograph mCT scanner (Siemens, Germany)

–

No

OSEM

–

Yes

Yes

4-mm full-width-at-half-maximum Gaussian

Manual

–

 > 3

[102]

19

440 ± 2.0

GE Discovery STE PET/CT Scanner, GE Discovery 600 PET/CT Scanner

–

Yes

OSEM

–

No

No

4.29 mm, 7 mm, or 10 mm full-width-at-half-maximum Gaussian filter

Automated

–

–

[123]

20

370

GE Discovery VCT scanner (Waukesha, WI)

–

No

2D PET: OSEM 3D PET: Iterative- Vue Point algorithm

2D: 128 × 128 3D: 256 × 256

No

No

3 mm, 5 mm, 6 mm post-filtration width

Semi-automated

–

–

[126]

21

229.4 ± 22.2

Biograph 64 mCT scanner (Siemens)

–

No

FBP and OSEM

256 × 256,128 × 128

Yes

Yes

2.5 mm, 3.5 mm, 4.5 mm, 5.5 mm full-width at half-maximum Gaussian

Semi-automated

–

 < 5

[127]

22

740

Siemens Biograph PET/CT scanner

21–40

Yes

OSEM

3D: 168 × 168 4D: 256 × 256

No

No

3D scan: 7 mm full-width at half-maximum Gaussian 4D scan: 5 mm full-width at half-maximum Gaussian

Automated

–

 < 3

[147]