Skip to main content

Table 2 Technique parameters AP pelvic radiography in our hospital (MUMC+)a

From: Gonad shielding in pelvic radiography: modern optimised X-ray systems might allow its discontinuation

Room

n

kVp

Tube current

texposure

Tube load

KAP

FID

KfiA

ESAK

   

mA

ms

mA.s

Gy cm2

cm

mGy

mGy

1

238

81 ± 1

806 ± 15

20 ± 13

16 ± 10

0.48 ± 0.29

126 ± 11

0.52 ± 0.31

0.78 ± 0.46

2

110

85 ± 0

472 ± 2

49 ± 28

23 ± 13

0.70 ± 0.41

142 ± 7

0.61 ± 0.36

0.91 ± 0.54

  1. a Inherent filtration X-ray tube 3 mm Al, added filtration 0.1 mm Cu, anode angle 16°, a 25-cm distance from skin on X-ray entrance side to image receptor is assumed, backscatter factor 1.49
  2. KAP kerma area product, FID ray focus to image receptor distance, KfiA kerma free in air at entrance position on skin (patient removed), ESAK entrance surface air kerma including backscatter (=KfiA × backscatter factor)