Skip to main content

Table 4 Impact of implant position and Eklund Manœuvre on the achievement of image quality criteria (p, p value; NS, not significant)

From: Study of breast implants mammography examinations for identification of suitable image quality criteria

Criteria

Implant position

Eklund Manœuvre

Eklund

No Eklund

Global

Subglandular implant

Subpectoral implant

Global

Breast centrally placed

p = < 0.01

NS

p = < 0.05

NS

NS

NS

Pectoral muscle (PM) visibility

p = < 0.001

p = < 0.001

p = < 0.001

NS

p = 0.002

NS

Visualisation of retroglandular adipose tissue

p = < 0.001

NS

p = < 0.001

p = < 0.001

p = < 0.001

p = < 0.001

Medial border of the breast included on the image

NS

p = 0.01

p = 0.05

p = < 0.001

NS

p = < 0.001

Axillary tail demonstrated

p = 0.01

NS

NS

p = < 0.001

NS

p = < 0.001

Nipple in the midline (+/− 10°)

NS

NS

NS

p = < 0.001

p = 0.01

p = < 0.001

Nipple in profile or transected by skin

NS

NS

NS

NS

NS

NS

No skin folds

p = < 0.001

NS

p = < 0.001

p = < 0.001

p = < 0.001

p = < 0.001

Spread of glandular tissue

NS

NS

p = 0.01

p = < 0.001

p = < 0.001

p = < 0.001

Sharpness

NS

NS

NS

NS

p = < 0.05

NS

Artefacts

p = 0.01

NS

NS

p = < 0.001

p = < 0.05

p = < 0.001

Symmetrical mirror images R/L images

p = 0.01

NS

p = < 0.001

p = < 0.001

p = 0.004

p = < 0.001

Beam penetration

NS

NS

NS

NS

NS

NS

Contrast

NS

NS

NS

NS

p = 0.002

p = < 0.05

Total of significant criteria

7

2

7

8

9

9