Skip to main content

Table 1 Imaging parameters

From: Clinical evaluation of left ventricular function and morphology using an accelerated k-t sensitivity encoding method in cardiovascular magnetic resonance

 

2D-SSFP

Accelerated k-t SENSE

ECG triggering

Retrospective

Prospective

TE/TR (ms)

1.7/3.4

1.7/3.4

FOV (mm)

320 × 360

320 × 360

Image matrix

192 × 192

176 × 192

Spatial resolution (mm)

1.67 × 1.87

1.82 × 1.87

Temporal resolution (ms)

49

37

Slice thickness/spacing (mm)

10/0

10/0

Flip angle (°)

60

60

Bandwidth (Hz/pixel)

977

1302

Cardiac phases (n)

25

32

View sharing (segments)

14

11

Breath-holds (n)

10–12

5–6

Slices per breath-hold

1

2

Acceleration factor

Parallel imaginga

4x

Acquisition duration (s)

161 ± 25

67 ± 14

  1. TE echo time, TR repetition time, FOV field of view, 2D-SSFP 2D steady-state free precession
  2. aParallel imaging intrinsic acceleration