Table 2 Royal London Hospital MRI protocol for imaging of craniocervical junction and cervical spine for ligamentous and neuroparenchymal traumatic injury evaluation (Siemens Avanto 1.5 Tesla MRI scanner)

From: The craniocervical junction: embryology, anatomy, biomechanics and imaging in blunt trauma

Sequence

TR (ms)

TE (ms)

FOV (mm)

NEX (averages)

Matrix

Slice thickness (mm)

Sag T2

3500

83

240

2

269 × 384

3/3.3

Sag T1 TSE

740

11

240

3

269 × 384

3/3.3

Sag STIR

4000

30

240

2

205 × 256

3/3.3

Ax T1 TSE

610

12

200

2

190 × 320

4/4.4

Ax T2

8250

91

220

2

196 × 320

4/4.4

T2 3DSPACE COR

1500

129

160

1.6

261 × 256

0.8