Skip to main content

Table 1 Comparison of mean CT doses between current and optimised protocols (confidence interval p < 0.05)

From: Visual grading characteristics and ordinal regression analysis during optimisation of CT head examinations

CT head

GE

BrightSpeed

  

Philips

Brilliance

  

Independent samples t-test

Current

Optimised

p

Current

Optimised

p

 

n =24

n =20

 

n =24

n =20

 

CTDIvol (mGy)

35.8

33.1

0.00

39.6

28.3

0.00

(Range)

(30.0-37.6)

(29.8-36.1)

 

(39.6-39.6)

(28.3-28.3)

 

DLP (mGy-cm)

489.4

461.5

0.03

694.8

637.3

0.03

(Range)

(420.0-564.0)

(389.2-575.1)

 

(491.0-820.0)

(541.1-823.1)