Table 4 Suggested IV contrast injection regiments

From: Imaging in Transcatheter Aortic Valve Replacement (TAVR): role of the radiologist

64-Detector scanner

320-Detector scanner

Gated cardiac CTA

Gated CTA chest

Bolus injection, 20 ml IV contrast

Automatic triggering

ROI: proximal descending aorta

ROI: proximal descending aorta

Peak enhancement selected

Threshold of +200 HU used

Main injection: 4 ml/s, 70–80 ml

Rate: 4 ml/s, 120–140 ml

Non-gated CTA abdomen, pelvis

Non-gated CTA abdomen, pelvis

Additional 50 ml is injected

Done immediately after chest CTA

Injection timing based on bolus results

No additional IV contrast required

Total:

Total:

130–150 ml

120–140 ml

  1. ROI Region of interest