Table 3 Recommended tests for QC for image detection and acquisition in mammographic systems (include testing the x-ray generation and image receptor)

From: Quality assurance and quality control in mammography: a review of available guidance worldwide

 

International

National

Regional

Frequency

X-ray generation

Test type

Target parameter to assess

EC (2006)

IAEA-SF (2009)

IAEADM (2011)

ACR (1999)

UK/IPEM (2005)

Ireland (2008)

Australia (2008)

RANZCR (2009)

UK/NHSBSP (2009)

Canada (Quebec) (2001/2006)

Total (10)a

X-ray source

Focal spot size

Y

N

N

Y

Y

Y

N

N

N

Y

5

Source-to-image distance

Y

N

N

N

Y

Y

N

N

Y

Y

5

Alignment of X-ray field/image receptor

Y

Y

Y

Y

Y

Y

Y

N

Y

Y

9

Radiation leakage

Y

Y

N

N

Y

Y

Y

N

Y

N

7

Tube output

Y

N

Y

Y

Y

Y

Y

N

Y

Y

8

Tube voltage and beam quality

Reproducibility and accuracy

Y

Y

Y

Y

Y

Y

Y

N

Y

Y

9

Half value layer (HVL)

Y

Y

Y

Y

Y

Y

Y

N

Y

Y

9

AEC system performance

Optical density control setting: central value and difference per step

Y

Y

N

Y

Y

N

Y

N

N

Y

7

Back-up timer and security cut-off

Y

N

N

N

Y

Y

Y

N

Y

N

5

Short term reproducibility

Y

N

N

Y

Y

Y

Y

Y

Y

R

8

Long term reproducibility

Y

N

Y

N

N

N

Y

N

N

R

4

Object thickness and tube voltage compensation

Y

Y

Y

Y

Y

Y

Y

Y

Y

Y

10

Correspondence between AEC sensors

Y

N

N

N

N

N

Y

Y

Y

Y

5

Compression

Compression force

Y

N

Y

Y

Y

N

Y

Y

N

Y

7

Compression plate alignment

Y

Y

Y

Y

N

Y

Y

Y

N

Y

8

Bucky and image receptor

Anti-scatter grid

Grid system factor

Y

N

N

N

Y

Y

N

N

N

Y

4

Grid imaging

Y

N

N

N

Y

Y

N

N

N

Y

4

Screen-film

Inter cassette sensitivity and attenuation variation and optical density range

Y

Y

N

Y

Y

N

Y

N

N

Y

6

Screen-film contact

Y

Y

N

Y

Y

N

Y

N

N

Y

6

Image receptor response (digital)

Response function

Y

N

Y

N

N

Y

Y

N

Y

R

6

Noise

Y

N

Y

N

R

Y

Y

N

Y

Y

7

Missed tissue at chest wall edge (digital

 

Y

N

Y

N

N

N

Y

N

Y

Y

5

Image receptor homogeneity and stability (digital)

Image receptor homogeneity

Y

N

Y

N

Y

Y

Y

Y

Y

Y

8

Detector element failure (DR systems)

Y

N

N

N

N

Y

Y

Y

N

N

4

Uncorrected defective detector elements (DR systems)

Y

N

N

N

N

Y

N

N

N

N

2

Inter-plate sensitivity variation (CR)

 

Y

N

Y

N

N

N

Y

Y

N

Y

5

Other sources of radiation (CR)

 

Y

N

N

N

N

N

N

N

N

Y

2

Fading of latent image (CR)

 

Y

N

N

N

N

N

N

N

Y

Y

4

System properties

Dosimetry

 

Y

Y

Y

Y

Y

Y

Y

N

Y

Y

9

Image quality

Spatial resolution

Y

Y

Y

Y

Y

Y

Y

N

Y

Y

9

Image contrast

Y

Y

Y

Y

R

Y

Y

Y

Y

Y

10

Threshold contrast visibility

Y

Y

N

Y

R

Y

Y

Y

Y

Y

9

Exposure time

Y

N

N

N

Y

Y

Y

N

N

R

5

Modulation transfer function and noise power spectrum (optional)

Y

N

Y

N

Y

Y

Y

N

Y

Y

7

Artefacts

Y

N

N

Y

Y

N

Y

Y

Y

Y

7

Geometric distortion and artefacts evaluation

Y

N

Y

Y

N

Y

Y

Y

Y

Y

8

Ghost image/erasure thoroughness

Y

N

Y

N

N

Y

Y

N

Y

Y

6

  1. Y yes/exists, N no/not provided, R referred without detail on the methodology
  2. aTotal refers to the total number of guidance documents that recommend the test