Skip to main content

Table 1 Dose results of CT protocols used in this national survey on CT radiation doses

From: A national survey on radiation dose in CT in The Netherlands

Protocol name

Clinical indication

Relative frequency (%)

Median scan range (mm)

Median (P75) DLP (mGy cm)

Median (P75) E103 (mSv)

Ratio max/min E

Brain

Haemorrhage

23.8

155

813.7 (935.6)

1.5 (1.8)

3.3

Sinus

Sinusitis

9.0

116

105.7 (133.4)

0.2 (0.3)

4.9

Neck

Standard

2.3

252

329.9 (404.3)

1.7 (2.1)

2.2

Neck-thorax-abdomen

Standard

0.7

880

985.1 (1,117.6)

13.8 (15.6)

1.7

Thorax

Standard

6.6

313

320.0 (346.5)

4.6 (5.0)

2.8

Thorax-liver

Lung cancer

4.1

401

542.0 (608.1)

8.1 (9.1)

2.7

Thorax-abdomen

Standard

9.5

687

813.8 (885.7)

11.6 (12.6)

3.9

Thorax HR

Interstitial Dz

3.1

273

142.8 (276.1)

2.1 (4.0)

13.4

Abdomen

Standard

20.0

441

567.0 (618.0)

8.0 (8.7)

4.8

Abdomen low-dose

Urolithiasis

2.1

399

301.6 (329.4)

4.5 (4.9)

3.3

CT urography

Haematuria

2.8

452

708.0 (981.6)

10.0 (13.8)

2.5

Liver multiphasic

Liver Tx

1.1

419

1,322.2 (1,496.6)

19.4 (22.4)

1.4

Pancreas multiphasic

Adeno ca

0.6

409

899.0 (1,000.0)

13.2 (14.7)

2.0

Kidney multiphasic

RCC

0.4

381

1,166.5 (1,371.2)

17.0 (20.2)

3.1

CT coronary angio

CAD

1.5

125

516.5 (671.4)

13.5 (20.1)

3.0

CT pulmonary angio

PE

4.8

279

291.5 (371.0)

4.2 (5.4)

2.7

CTA total aorta

Dissection

0.7

666

822.5 (836.8)

11.7 (11.9)

1.1

CTA abdominal aorta

AAA

1.8

474

555.5 (727.1)

7.8 (10.3)

3.8

C-Spine

Standard

2.5

213

276.0 (320.9)

1.4 (1.6)

5.4

L-Spine

Standard

1.5

171

308.2 (405.5)

4.3 (5.7)

3.2

Bony pelvis

Standard

1.0

223

332.0 (382.1)

4.3 (4.9)

1.6

  1. P75 75th percentile, DLP dose length product, E 103 effective dose according to ICRP-103, Dz disease, Tx transplantation, ca carcinoma, RCC renal cell cancer, CAD coronary artery disease, PE pulmonary embolism, AAA abdominal aortic aneurysm