Table 2 Mean reconstruction speed for filtered back projection and iterative CT reconstructions

From: Iterative reconstruction does not substantially delay CT imaging in an emergency setting

 

Speed (slices/s)

P valuea

Slices (N)

 

FBP

L2

Percentual delay

L4

Percentual delay

FBP vs L2

FBP vs L4

L2 vs L4

 

CTA b

26.7 ± 1.1

18.7 ± 0.5

−30 %

18.0 ± 0.6

−33 %

<0.001

<0.001

0.040

647.8 ± 16.6

 

FBP

L1

Percentual delay

L6

Percentual delay

FBP vs L1

FBP vs L6

L1 vs L6

 

Total body c

11.2 ± 0.3

7.1 ± 0.1

−37 %

7.3 ± 0.2

−35 %

<0.001

<0.001

0.237

200.2 ± 6.6

Abdominal c

20.7 ± 0.6

15.3 ± 0.3

−26 %

14.9 ± 0.3

−28 %

<0.001

<0.001

0.103

674.1 ± 21.8

Chest c

19.9 ± 0.6

14.0 ± 0.3

−30 %

14.2 ± 0.3

−29 %

<0.001

<0.001

0.475

389.6 ± 16.3

Cervical c

30.0 ± 0.3

15.5 ± 0.5

−48 %

15.4 ± 0.4

−49 %

<0.001

<0.001

0.683

476.2 ± 16.1

Total trauma CT

20.1 ± 1.6

13.2 ± 0.8

−34 %

13.2 ± 0.6

−34 %

<0.001

<0.001

0.753

1740.1 ± 213.5

  1. FBP filtered back projection, L2 iDose4 level 2, L4 iDose4 level 4
  2. aBased on dependent t-test
  3. bn = 25
  4. cn = 15