Skip to main content

Table 1 Mean reconstruction time for filtered back projection and iterative CT reconstructions

From: Iterative reconstruction does not substantially delay CT imaging in an emergency setting

 

Time (s)

P valuea

Slices (N)

 

FBP

L2

Percentual delay

L4

Percentual delay

FBP vs L2

FBP vs L4

L2 vs L4

 

CTA b

25.5 ± 1.4

35.6 ± 1.8

+40 %

37.6 ± 2.4

+47 %

<0.001

<0.001

0.057

647.8 ± 16.6

 

FBP

L1

Percentual delay

L6

Percentual delay

FBP vs L1

FBP vs L6

L1 vs L6

 

Total body c

18.1 ± 0.9

28.3 ± 1.0

+56 %

27.8 ± 1.1

+54 %

<0.001

<0.001

0.255

200.2 ± 6.6

Abdominal c

33.2 ± 1.7

44.3 ± 1.8

+33 %

45.7 ± 2.1

+38 %

<0.001

<0.001

0.139

674.1 ± 21.8

Chest c

20.1 ± 1.1

28.2 ± 1.4

+40 %

27.8 ± 1.5

+38 %

<0.001

<0.001

0.633

389.6 ± 16.3

Cervical c

15.9 ± 0.6

30.8 ± 0.9

+94 %

31.0 ± 0.9

+95 %

<0.001

<0.001

0.780

476.2 ± 16.1

Total trauma CT

87.3 ± 14.6

131.7 ± 16.7

+51 %

132.2 ± 17.9

+51 %

<0.001

<0.001

0.921

1740.1 ± 213.5

  1. FBP filtered back projection, L2 iDose4 level 2, L4 iDose4 level 4
  2. aBased on dependent t-test
  3. bn = 25
  4. cn = 15