Skip to main content

Table 2 Imaging technique settings in AP pelvic radiographya

From: Gonad shielding in paediatric pelvic radiography: disadvantages prevail over benefit

Age

n

Filterb

High voltage

mAs valuec

AEC

Grid

(years)

(mm Cu)

(kV)

(mAs)

(yes/no)

(yes/no)

0-1

36

-

50

4.2

n

n

1-5

63

0.1

55

5.2

n

n

5-10

43

0.1

55/60

6.5

n

n

10-15d

27

0.1

70

12

y/n

y/n

10-15e

26

0.1

81

13

y

y

  1. aThe average focus-image receptor distance was 117 ± 24 cm
  2. bInherent filtration was 3.3 mm Al-equivalent
  3. cTube current exposure time product
  4. dSmaller children, typically no grid and no AEC
  5. eLarger children, typically with grid and/or AEC