Skip to main content

Table 1 Image parameters of the 4 sequences of NCMRA

From: 3D whole-heart noncontrast coronary MR angiography based on compressed SENSE technology: a comparative study of conventional SENSE sequence and coronary computed tomography angiography

Sequence type

SENSE

CS4

CS5

CS6

FOV (mm)

350 × 350

350 × 350

350 × 350

350 × 350

Slice thickness (mm)

1.5

1.5

1.5

1.5

Slice number

140

140

140

140

Reconstruction voxel size (mm)

0.63 × 0.63 × 0.75

0.63 × 0.63 × 0.75

0.63 × 0.63 × 0.75

0.63 × 0.63 × 0.75

Acceleration factors

SENSE(2)

CS (4)

CS (5)

CS (6)

TR (ms)

3.3

3.3

3.3

3.3

TE (ms)

1.3

1.3

1.3

1.3

Flip angle (0)

17

17

17

17

Scan time (s)

343.27 ± 44.57

269.91 ± 34.82

214.97 ± 28.31

190.19 ± 21.41