Skip to main content

Table 4 Mean effective dose for the original scan (100%) and the simulated dose reductions (70%, 50%, 30%) for boys and girls, the respective scan region (hip, knee, ankle) and the sum of the effective dose (mean)

From: Ultra-low-dose computed tomography for torsion measurements of the lower extremities in children and adolescents

Dose

Mean effective dose hip boys (mSv)

Mean effective dose hip girls (mSv)

Mean effective dose knee boys (mSv)

Mean effective dose knee girls (mSv)

Mean effective dose ankle boys (mSv)

Mean effective dose ankle girls (mSv)

Sum of effective dose boys (mean) (mSv)

Sum of effective dose girls (mean) (mSv)

100%SD

0.420 ± 0.172

0.450 ± 0.173

0.030 ± 0.006

0.030 ± 0.005

0.010 ± 0.002

0.010 ± 0.003

0.460 ± 0.177

0.490 ± 0.178

70%

0.294

0.315

0.021

0.021

0.007

0.007

0.322

0.343

50%

0.210

0.225

0.015

0.015

0.005

0.005

0.230

0.245

30%

0.126

0.135

0.009

0.009

0.003

0.003

0.138

0.147

  1. SD, standard deviation