Skip to main content

Table 3 Analysis by examination type of average CTDIvol per sequence and total DLP per examination, for distribution of median values per CT scanner

From: Computed tomography radiation doses for common computed tomography examinations: a nationwide dose survey in United Arab Emirates

CT protocol

Median values of CTDIvol per sequence

Median values of DLP per examination

Range, min–max (max/min ratio)

1st quartile (25%)

Median (50%)

Mean

3rd quartile (75%)

Range, min–max (max/min ratio)

1st quartile (25%)

Median (50%)

Mean

3rd quartile (75%)

Head without CM

6.4–54.7 (8.6)

27.2

29.9

32.1

39.6

384.8–743.0 (1.9)

491.4

639.3

592.7

693.1

Head with CM

25.0–54.7 (2.2)

34.6

41.4

41.1

48.0

430.0–1168.0 (2.7)

676.4

772.2

780.1

818.7

Abdomen-pelvis without CM

4.4–15.7 (3.6)

9.1

11.0

10.9

13.5

330.0–1186.8 (3.6)

551.1

656.2

683.2

811.8

Abdomen-pelvis with CM

11.3–25.0 (2.2)

13.3

16.9

17.2

20.4

363.0–1698.0 (4.7)

533.6

606.5

825.9

1023.1

Chest

4.7–12.3 (2.6)

5.6

7.1

7.8

10.2

137.5–360.8 (2.6)

227.2

251.7

255.0

276.2