Table 1 Reference values of native myocardial T1 mapping at different studies

From: T1 mapping: characterisation of myocardial interstitial space

  Field strength Scanner Reception coil T1 mapping sequence T1 values (ms)
Messroghli et al. [42] 2006 1.5 T Gyroscan Intera CV, Philips MOLLI 980 ± 53
Piechnik et al. [36] 2010 1.5 T Avanto, Siemens 32-channel ShMOLLI 966 ± 48
  1.5 T Avanto, Siemens 32-channel MOLLI 976 ± 46
  3 T Trio, Siemens 16-channel ShMOLLI 1,166 ± 60
  3 T Trio, Siemens 16-channel MOLLI 1,169 ± 45
Lee et al. [58] 2011 3 T Verio, Siemens 32-channel MOLLI 1,315 ± 39
Rogers et al. [59] 2013 1.5 T Philips, Best 32-channel MOLLI 976 ± 37
  3 T Philips, Best 32-channel MOLLI 1,108 ± 67
Von Knobelsdorff et al. [49] 2013 3 T Verio, Siemens 32-channel MOLLI 1,158 (range, 1,005–1,295)
Kellman et al. [82] 2013 1.5 T Avanto, Siemens 32-channel MOLLI 1,012 ± 25
Piechnik et al. [83] 2013 1.5 T Avanto, Siemens 16-channel ShMOLLI 962 ± 25
  1.5 T Avanto, Siemens 32-channel ShMOLLI 962 ± 25
Puntmann et al. [74] 2013 3 T Achieva TX, Philips 32-channel MOLLI 1,070 ± 55
Fontana et al. [76] 2014 1.5 T Avanto, Siemens 16-channel ShMOLLI 967 ± 34
Liu et al. [84] 2014 3 T Trio, Siemens 12-channel MOLLI 1,232 ± 51