Skip to main content

Table 1 Technical parameters of a vascular ring protocol

From: Computed tomography in the evaluation of vascular rings and slings

Parameters

ECG gated

Non-ECG gated

Gating

Prospective ECG triggered (40 % R-R interval)

None

Scan mode

Axial, step and shoot

Helical

Detector collimation

128 × 0.625 mm, adaptive z collimation

128 × 0.625 mm, adaptive z collimation

Gantry rotation time

270 ms

270 ms

Temporal resolution

135 ms

135 ms

Scan length

Thoracic inlet to diaphragm

Thoracic inlet to diaphragm

Pitch

1.375

Tube voltage

80 kV

80 kV

Tube current

0–3 kg–60 mAs/slice

3–6 kg–80 mAs/slice

6–10 kg–100 mAs/slice

10–15 kg–120 mAs/slice

0–3 kg–60 mAs/slice

3–6 kg–80 mAs/slice

6–10 kg–100 mAs/slice

10–15 kg–120 mAs/slice

Reconstructed slice thickness

0.9 mm

0.9 mm

Slice interval

0.45 mm

0.4 mm

Reconstruction algorithm 4

Idose- level 3

Idose- level 3

Contrast

1–2 ml/kg

1–2 ml/kg

Injection rate

1–3 ml/s

1–3 ml/s

Contrast timing

Empirical

Empirical