Skip to main content

Table 2 Median and interquartile range of DLP values in unenhanced and enhanced CT examinations before and after radiological staff training (involving both technologists and radiologists). DLP values are expressed in mGy∙cm

From: Real practice radiation dose and dosimetric impact of radiological staff training in body CT examinations

 

Before training (centre 4)

After training (centre 4)

p value

Median

IQR

No

Median

IQR

No

Chest (C−)

680

513÷1053

127

413

323÷577

100

<0.001

Chest (C+)

1253

820÷1767

35

560

380÷760

50

<0.001

Abdomen-pelvis (C−)

967

633÷1247

191

607

460÷793

50

<0.001

Abdomen-pelvis (C+)

3227

1960÷4547

79

1760

1140÷2580

50

<0.001

Whole body (C−)

1167

733÷1600

36

800

547÷1113

50

<0.001

Whole body (C+)

2180

1373÷3400

509

1479

1053÷2180

106

<0.001