Skip to main content

Table 1 Different GRE sequences with fixed TR and variable TE and flip angle, for MRI with fields of 0.5, 1, and 1.5 Tesla, used in the method of Gandon et al from the University of Rennes, France (see [24, 29])

From: Quantification of iron concentration in the liver by MRI

0.5 T

1.0 T

1.5 T

TR

TE

Flip

TR

TE

Flip

TR

TE

Flip

120

14

90°

120

7

90°

120

4

90°

120

14

20°

120

7

20°

120

4

20°

120

28

20°

120

14

20°

120

9

20°

   

120

21

20°

120

14

20°

      

120

21

20°